Redline Operating Manual RE / RF / RI 12/2010
page 21/46
•
Permissible ambient humidity: 70 % r.H. max., non-condensing.
•
Installation height red LINE: max. 2000 m /
6562 ft
above sea level.
•
Installation height redline-UL: max. 5000 m /
16,404 ft
above sea level.
Maintain a minimum distance of 250 mm /
9.84 in
between each unit. Wall distances: rear 100 mm /
3.94
in
, sides 160 mm /
6.30 in
.
To completely separate the unit from the power supply, you must disconnect the power plug. Install the
unit in a way that the power plug is easily accessible and can be easily pulled in case of danger.
Do not install or operate the unit in hazardous locations.
DANGER
Explosion hazard.
Danger of death.
∅
Do NOT operate redLINE chambers in potentially explosive areas.
¾
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit.
redLINE chambers are not waterproof. Do not install the chamber in locations where water could be
sprayed about, e.g. in the vicinity of sinks or safety showers.
DANGER
Electrical hazard.
Danger of death.
¾
redLINE chambers must NOT become wet during operation or maintenance work
3.5
Stacking
Two redLINE chambers can be piled on top of each other. To ensure that they are safely positioned, both
devices should be attached to one another using the flat assembly bars supplied. One flat assembly bar
is supplied with each redLINE chamber.
CAUTION
Sliding or tilting of the upper chamber.
Damage to the chambers.
∅
Never stack more than two redLINE chambers.
¾
Always secure stacked redLINE chambers using both flat assembly bars supplied