BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 03/2022
page 26/60
11.1 Cleaning
Disconnect the chamber from the power supply before cleaning. Disconnect the power plug.
The interior of the chamber must be kept clean. Thoroughly remove any residues of the
charging material
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Exterior surfaces
inner chamber
racks
door gaskets
Standard commercial cleaning detergents free from acid or halides.
Alcohol-based solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Instrument panel
Standard commercial cleaning detergents free from acid or halides.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Zinc coated hinge parts
rear chamber wall
Standard commercial cleaning detergents free from acid or halides.
Do NOT use a neutral cleaning agent on zinc coated surfaces.
Do not use cleaning agents that may cause a hazard due to reaction with components of the device or the
charging material. If there is doubt regarding the suitability of cleaning products, please contact BINDER
service.
We recommend using the neutral cleaning agent Art. No. 1002-0016 for a thorough cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
Any corrosive damage caused by a lack of cleaning, is excluded from liability by BINDER
GmbH.
NOTICE
Danger of corrosion by using unsuitable cleaners.
Damage to the chamber.
∅
Do NOT use acidic or chlorine cleaning detergents.
∅
Do NOT use a neutral cleaning agent on other kind of surfaces e.g., the zinc coated
hinge parts or the rear chamber wall.
For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove cleaning agents from the surfaces with a moistened towel.
Let the chamber dry.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
With every decontamination method, always use adequate personal safety controls.
Following cleaning, leave the chamber door open or remove the access port plugs.
The neutral cleaning agent may cause health problems in contact with skin and if ingested.
Follow the operating instructions and safety hints labeled on the bottle of the neutral clean-
ing agent.
Содержание 9090-0014
Страница 39: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 39 60 14 7 2 BD S BD S UL 115 mm...
Страница 40: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 40 60 14 7 3 ED S ED S UL 56 mm...
Страница 41: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 41 60 14 7 4 ED S ED S UL 115 mm...
Страница 42: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 42 60 14 7 5 FD S FD S UL 56 mm...
Страница 43: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 43 60 14 7 6 FD S FD S UL 115 mm...
Страница 45: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 45 60...
Страница 46: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 46 60 15 2 EU Declaration of Conformity for ED S...
Страница 47: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 47 60...
Страница 48: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 48 60 15 3 EU Declaration of Conformity for FD S...
Страница 49: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 49 60...
Страница 51: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 51 60...
Страница 52: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 52 60...
Страница 53: ...BD S BD S UL ED S ED S UL FD S FD S UL E1 03 2022 page 53 60...
Страница 61: ......