KBF-S / KBF-S-UL (E6) 11/2021
page 35/95
If the function “Language selection at restart” has been activated (chap. 11.5, factory setting ON), the fol-
lowing settings are checked upon start up:
•
Menu language
(chap. 11.1):
Use the
arrow buttons
to select the desired language, confirm with the
OK button
•
Temperature unit
(chap. 11.2):
Use the
arrow buttons
to select the desired temperature unit, confirm with the
OK button.
•
Current date
(chap. 11.3), format DD MM YYYY:
Use the
arrow buttons
to set the day, continue with the
OK button
.
Use the
arrow buttons
to set the month, continue with the
OK button
.
Use the
arrow buttons
to set the year, confirm with the
OK button
•
Current time
(chap. 11.4), format HH:MM:
Use the
arrow buttons
to set the hours, continue with the
OK button
.
Use the
arrow buttons
to set the minutes, confirm with the
OK button
7. Temperature and humidity set-point entry
Required access level: “User”.
Setting ranges
Control ranges
Temperature
-5 °C / 23
°F
up to 70 °C /
158 °F.
0 ºC /
32 °F
up to 70 °C /
158 °F
without humidity
10 °C /
50 °F
up to 70 °C /
158 °F
with humidity
Humidity
20 % r.h. up to 80 % r.h.
20 % r.h. to 80 % r.h.
See climatic diagram, chap. 19.
For the control range of temperature and relative humidity, see the temp. / humidity diagram chap. 19).
With safety controller mode “
Limit
”, adapt the safety controller always when you changed the
temperature set-point. Set the safety controller value by approx. 2 °C to 5 °C above the tem-
perature set-point (chap. 10.2).
Recommended setting: safety controller mode “
Offset
” with safety controller value 2 °C.
When operating without humidity by activated function “Humidity off” (chap. 8), set the humid-
ity tolerance range to “0” in order to avoid tolerance range alarms (chap. 12).
7.1 Temperature set-point entry
Path:
Normal display Setpoints
Temperature
Press the
OK button
to enable the setting.
Temperature setting.
The current setting flashes. Enter the desired set-point with the
arrow
buttons
.
Confirm the entry with the
OK button
.
Temperature
With the
arrow-down button
you can continue with the humidity set-point entry (chap. 7.2).
With the
Back button
you can go back to the “
Setpoints
” submenu and, repeatedly pressing it, to
Normal
display
.
Содержание 9020-0366
Страница 81: ...KBF S KBF S UL E6 11 2021 page 81 95 25 7 Dimensions Dimensions size 115...
Страница 82: ...KBF S KBF S UL E6 11 2021 page 82 95 Dimensions size 240...
Страница 83: ...KBF S KBF S UL E6 11 2021 page 83 95 Dimensions size 720...
Страница 84: ...KBF S KBF S UL E6 11 2021 page 84 95 Dimensions size 1020...
Страница 86: ...KBF S KBF S UL E6 11 2021 page 86 95...
Страница 87: ...KBF S KBF S UL E6 11 2021 page 87 95...
Страница 89: ...KBF S KBF S UL E6 11 2021 page 89 95...