12. Notification and alarm functions
12.1 Notification and alarm system overview (auto diagnosis system)
The alternating climate chamber MK provides notification and alarm functions.
•
Visual indications of notifications or error message are blue notes on the display of the MB1 controller.
•
Visual indications of an alarm message are red notes with an alarm bell symbol.
In addition, there is an audible alert, if you did not deactivate the buzzer in the “User level” menu (chap.
Event
Note
(blue field)
Alarm
(red field)
Door open
DOOR OPEN
immediately
Operation line 1 (bedew protection) activated
DRY
immediately
Limit value of safety controller exceeded
TEMPERATURE LIMIT
immediately
Power failure
With option over/under temperature safety device class 2 (chap. 11.3):
Exceeding the maximum / minimum tempera-
ture
TEMP CLASS
immediately
With option keyboard locking (chap. 14.3):
Keyboard locked
KEY LOCK
immediately
The indicated intervals refer to the time after occurrence of the error or notified condition.
12.2 Resetting the notifications or alarm messages
The “RESET” button, which serves to acknowledge and reset the indication, will become visible automati-
cally whenever a notification or an alarm message appears.
1.
Depending on the type of error, remove the cause of the disturbance or wait until the unit compensates
for the reason of the error.
2.
Press the “RESET” button to reset the notification or alarm message.
CAUTION
In case the “RESET” button does not cancel the notification or alarm indication, the reason
for the disturbance was not removed correctly
Contact BINDER service.