Binatone Babysebse 5 Скачать руководство пользователя страница 1

 Breathing & Movement Monitor 

5

Medically approved

Sensor | Do not use | Sensor

Slot

I

llustration 6

Sensor | Do not use | Sensor

Hanging hook

Blue breathing 

movement 

indicator light

Red alert 

alarm light

Low battery 

warning light

On/Off

button

I

llustration 8

Speaker 

Control

Unit

Cable

Cable

Bed/Crib 

base

Control unit holder

I

llustration 5

1

3

2

Ensure you read this user guide before installing and 

using the product.

Babysense is a high quality, high sensitivity electronic 

device which must be handled with care: 

• Do not handle roughly, drop or throw.

• Follow the installation instructions carefully to 

ensure that it is installed securely and correctly.

• Babysense is not a toy. Do not allow children of any

 

age to play with it.

• Do not put any additional baby or pet in the bed.

Babysense cannot be used for two infants in one bed. 

• You must undertake the performance test daily of 

the Babysense monitor (as set out on page 

3).

• Do not operate a vaporiser or ventilator directly 

towards the bed.

• To avoid possible danger from sensor cords, 

ensure that excess cords are pulled back under 

the mattress, secured with the included cable tie 

and are not visible or reachable.

• 

Install 4 new and identical size AA 1.5 volt batteries, 

preferably alkaline. 

• 

IMPORTANT: 

It is strongly advised that every 

caregiver learn how to give first aid.

"Babysense"

 monitors and alerts, but cannot prevent 

cessation of breathing. The responsibility for applying first 

aid rests on caretakers. The  "Babysense" senses breathing

 

movements and motion only, and cannot monitor 

other occurrences. The purpose of the device is to alert

 

when stopping or slowing of breathing and 

movement occur.

 

"Babysense"

 is a unique safety system which has helped 

save infants, giving parents (or caretaker) assurance, 

by enabling intervention in the event of irregular 

breathing movement or distress, before it becomes 

critical. There are a number of cautionary steps 

which are internationally recommended for 

SAFE 

SLEEPING

* Always lay your baby to sleep on his or her back.

* Refrain from smoking near or around your baby.

* Avoid over-heating the baby's surroundings.

* Do not leave any objects near your baby that can 

cause suffocation. 

 

Babysense is an addition to these precautions. 

 

"Babysense" 

is a home breathing movement monitor 

for basically healthy infants up to one year of age, 

and is not to be relied on for use with high-risk infants 

requiring constant medical supervision.

"Babysense"

 continually detects the baby's motion and 

breathing movements, and will set off a sound and 

visual alarm if movement stops for 20 seconds or if the 

breathing rate slows to below 10 breaths per minute  if 

the baby's breathing. 

The system's non-touch sensor 

pads under the mattress monitor baby's movement 

without restraining him.

1. 

Insert, in accordance with the markings, 4 new and 

identical size AA 1.5 volt batteries, preferable alkaline. 

Close the battery cover.

Note:

 Do not mix old and new batteries.

Note

: For replacing the batteries, remove the battery 

cover by inserting a flat screwdriver into the opening 

recess (see illustration 7).

Note

: Please dispose of used batteries in accordance 

with local waste regulations. 

2. 

Place control unit holder on the base of the crib 

or bed and in between two bars as shown in 

illustration 1.

 

3. Make sure the sensitivity adjustor, located on the back

 

of the sensor pads, is positioned on level 3 

(see illustration 2).

4.

 Sensor pad installation:

 

• When using for an infant not yet able to crawl, place 

one sensor pad as shown in illustration 3. Make sure the 

sensor is placed underneath the baby's position in the crib.

Warnings

Principal Of Operation 

Installation

Please note

9.

 

Place mattress firmly on top of sensor pads. If the 

mattress cannot be in contact with the sensor pads, 

or if the bed base is soft, padded or contains 

springs, place extra plywood on the bed base, place 

the sensor pads on the plywood and place the 

mattress firmly on top. 

IMPORTANT

: To avoid possible danger from sensor

 

cords, ensure that excess cords are pulled back 

under the mattress, secured with the included cable 

tie and are not visible or reachable.

See illustration 8 for layout of control unit featuring:

• On/Off button. 

• Blue breathing movement indicator light. 

• 

Acoustic and visual red alarm light. 

• Low battery red warning light. 

"Babysense"

 is ready for operation and use once 

the installation is complete. To operate, follow the 

below instructions:

• Place your baby in the crib/bed.

• Turn control unit on by pressing On/Off button.

• Red control lights switch on briefly and a double 

click sound is heard. 

• When breathing and movement is sensed, the 

blue light flashes. 

• Acoustic any visual red light alarms are activated 

if the baby's breathing stops or slows significantly.

• If the red low-battery warning light flashes, 

replace all batteries as soon as possible. If in 

addition clicking is heard, all batteries must be 

replaced immediately. 

• Turn off by pressing On/Off button before taking 

your baby out of the crib/bed. 

User Guide

0

1

2

3

Illustration 2

• When the baby begins to crawl, and is able to move 

onto areas not "covered" by a single sensor pad, attach 

the second sensor and place the two sensors  as shown 

in illustration 3.  

PLEASE NOTE: 

The sensors shall be placed with the 

marking “THIS SIDE UP” facing upfront. 

5. 

Pass the cord(s) of the sensor pad(s) through 

special openings in the holder. (see illustration 4)

6. Attach end(s) of the cord(s) to the outlets

marked "SENSOR" on the back of control unit

(see illustration 5 and 6). When using one sensor

pad, attach to either “SENSOR” outlet.

PLEASE NOTE:

 Do not attach any cable to the 

blocked socket, marked "DO NOT USE"

7. Should the design of the crib/bed not allow proper 

installation with the control unit holder, please use the 

optional hanging hook (see illustration 6) and hang the 

control unit from the bottom edge of the crib.

 

8. Press control unit holder or hanging hook firmly 

into slot in back of control unit (see illustration 5 and 6).

"Babysense"

 is a highly  sensitive device which 

monitors baby's breathing movements through the 

mattress during sleep. There are two main causes of 

malfunction - technical malfunctions and 

malfunctions due to environmental factors.

 

In order to minimise technical problems, follow 

installation and operation instructions carefully.

Due to the high sensitivity of 

"Babysense"

, it can 

be influenced by fans, wind, air conditioning, and 

so forth. Such environmental effects could be 

picked up by the sensors on a similar level to that of 

breathing movements, thus preventing or delaying 

alarm activation.

 

Do not conduct performance test by picking up 

baby out of bed and waiting for the alarm to sound 

• The performance test should be conducted 

while the baby is not in bed. 

• Turn device on and take two steps back from bed.

 

• Ensure the alarm goes off within 20 seconds 

(due to absence of breathing movements).

• The alarm may sound with or without 

preliminary ticking. 

• Check the alarm can be heard in all locations 

throughout  the home or place of care.

• In case the alarm does not sound within 20 

seconds refer to troubleshooting at page 5.

• The performance test must be conducted daily.

 
 

Isolated cases of false alarms are possible. Check if caused 

by faulty connection of the sensor pad(s) to the control 

unit, connection of only one sensor pad when the baby 

can move off the immediate area, connection of two  

sensor pads far apart from each other when the baby is 

not crawling yet, or failure of the sensor pads to be 

directly against the mattress.

 

In case of false alarms, kindly refer to troubleshooting 

guide. If the problem cannot be solved, please 

contact your nearest 

BABYSENSE

 representative.

 
Act immediately according to the following: 

1. 

Approach the bed and check that baby is breathing. 

Note pallor (color of skin), muscle tone and movement.

2. 

If baby does not seem normal, pinch baby's heel 

and check that there is no cause for choking, 

such as an object lodged in baby’s throat. 

When breathing resumes, consult a doctor.

3. 

If breathing does not resume, apply first aid and seek 

medical assistance immediately.

Sensor | Do not use | Sensor

Slot

Battery 

cover 

Blocked socket

DO NOT USE

“Sensor”

socket

I

llustration 7

Opening 

recess

Speaker 

Layout of control unit

Operation

Testing Operation

Technical malfunction

Environmentally caused malfunction

Conduct this performance test daily

False alarms

In the event of alarm

Cables

I

llustration 4

Control unit 

holder

I

llustration 1

Control unit holder

Cables

Sensor

Sensor

Sensor 

pads

(Under mattress)

Control unit

I

llustration 3

Top crib/bed view

For babies not 

yet able to crawl

For crawling babies

Top crib/bed view

Отзывы: