Weight indication
Within the preset upper and lower limits.
Weight indication
Above the preset upper limit.
4.2 Weighing function keys
Refer to point 2 of Figure 4.1 on page 1-29 :
Zero-set weight
pressing this key resets the weight indication only
if the following condition are satisfied:
✔
the weight value must be within the -1% ÷ +3%
range of the weighing capacity for terminals
subject to legal verification or ± 50% for other
terminals;
✔
the weight must be stable;
✔
no tare must be entered.
Enter/cancel tare
On pressing this key, the weight on the scale is
acquired as the tare value, provided the following
conditions are satisfied:
✔
the weight must be stable;
✔
the weight must have a positive value;
✔
the weight must not exceed the maximum
capacity.
The display illuminates the symbols
and
.
On MD terminals, the weight indication will be
displayed using the division of the lower range. On
ME terminals, the net weight will be displayed
using the division of the range in within which it
falls.
D410
1-31
Содержание D410
Страница 2: ......
Страница 36: ...D410 1 36 ...
Страница 56: ...D410 1 56 ...
Страница 69: ...Key 1 Comx1 2 Comx2 Figure 6 8 Serial port designation log0129 jpg D410 1 69 ...
Страница 83: ......