2.7
Remote commands for digital cells
Hereinafter, in addition to the symbols previously described in
the following conventions will
apply:
c is the number that identifies the digital cell;
n and m are the number fields with spaces in front, decimal separator, - sign and + sign.
2.7.1
Cell point request
DPc<CR>
The terminal responds with the string:
n<CR><LF>
n = number of points (0 ÷ 200000)
2.7.2
Cell temperature request
DTc<CR>
The terminal responds with the string:
n<CR><LF>
n = cell temperature in °C (- 40.0 ÷ 100.0)
2.7.3
Cell software version and release request
DVc<CR>
The terminal responds with the string:
n<SP>m<CR><LF>
n = cell software version
m = cell software release
2.7.4
Power supply request
DAc<CR>
The terminal responds with the string:
n<SP>m<CR><LF>
n = cell feed voltage
m = strain gauge feed voltage
2.7.5
Cell serial number numbers request
DMc<CR>
The terminal responds with the string:
n<SP>m<CR><LF>
n = cell serial number registered on the cell
m = cell serial number registered on the terminal
2.7.6
Angle calibration coefficient request
DCc<CR>
The terminal responds with the string:
n<SP>m<CR><LF>
n = angle calibration coefficient loaded on the cell
m = angle calibration coefficient loaded on the terminal
in this case n and m can also take on an exponential format.