Dental Handpiece Maintenance Device 

Distributed by: Bien Air Dental SA 

Manufactured by: J.MORITA MFG. CORP.  

2021-05-21  Pub. No.: K345-91001-503 (EN)  Printed in Japan

Lubricare 2

INSTRUCTIONS FOR USE

Содержание Lubricare 2 HIM-2

Страница 1: ...Dental Handpiece Maintenance Device Distributed by Bien Air Dental SA Manufactured by J MORITA MFG CORP 2021 05 21 Pub No K345 91001 503 EN Printed in Japan Lubricare 2 INSTRUCTIONS FOR USE...

Страница 2: ...r cente IT ES RU Prima di utilizzare il prodotto leggere at tentamente le istruzioni per l uso possibile scaricarle su internet www bienair com support download center Nel caso si riscontrino diffico...

Страница 3: ...s manual thoroughly before using the device and pay close attention to warnings and cautions Keep this manual in a readily accessible place for quick and easy reference Trademarks and Registered Trade...

Страница 4: ...allation 12 6 1 2 Air Tube 13 6 1 3 Power Cord 14 6 2 Spray Cans 15 6 2 1 Spray Stands 16 6 2 2 Spray Cans 17 6 2 3 Spray Stand Adjustment 18 6 3 Couplings 19 6 4 Oil Absorbent Sheet 20 6 5 Oil Collec...

Страница 5: ...r 32 8 2 2 Air Regulator 32 8 3 Parts Replacement 33 8 3 1 Oil Absorbent Sheet 33 8 3 2 Oil Absorbent Pad 33 8 3 3 Door Oil Absorbent Sheet 33 8 3 4 Nozzle Skirt 34 8 3 5 Coupling O rings 34 8 3 6 Spr...

Страница 6: ...fessionals Do not use the Lubricare 2 for purposes other than lubrication and flushing as part of handpiece maintenance The term flushing as used in this manual means ejecting and removing cutting deb...

Страница 7: ...d trained technician authorized by the manufacturer For customers who use the Lubricare 2 in the EU If any serious incident occurs in relation to the device report it to a competent authority of your...

Страница 8: ...tal SA OF FICE To prevent infection always wear appropriate personal protective equipment such as surgical gloves when using or cleaning the device These precautions are especially critical for safe o...

Страница 9: ...d Lubrication Packing Inspection Maintenance Procedure Sterilization Cleaning and Lubrication The term cleaning used in this manual refers to an internal cleaning of the handpiece with SPRAYNET that i...

Страница 10: ...Absorbent Pad Qty 1 Oil Absorbent Sheet Qty 1 Spray Stands Qty 2 Power Cord Qty 1 Air Tube Qty 1 Door Oil Absorbent Sheet Qty 1 Cord type and plug shape depend on your country or region Front Door Sh...

Страница 11: ...r View Air Regulator Rear Cover R Rear Cover L Spray Stand Power Cord Front View Couplings sold sepa rately Chuck Lubri cation Unit Exhaust Opening Front Door Door Hinge Door Oil Absorbent Sheet Botto...

Страница 12: ...xcess oil inside the hand piece when the maintenance is completed Start Stop Switch Mode 1 Selection Switch Press this switch to start or stop cleaning or lubrication Press this switch to select Mode...

Страница 13: ...handpiece lubrica tion is in progress Line Selection Lamps Each lamp lights up when the corre sponding line is selected When perform ing the lubrication the corresponding line lamp will blink When th...

Страница 14: ...a well ventilated place and ventilate it regularly 6 1 Main Unit 6 Setup Install the main unit with at least the clearance shown below provided around the unit Place the device on a level and stable s...

Страница 15: ...the air tube a light tug to make sure it is properly connected Then supply air by opening the main air valve Do not give the air tube a hard pull Otherwise the unit may tip over Make sure that the air...

Страница 16: ...Power Cord Plug Power Cord Connector Power Cord Hole Power Cord Connector 4 Pull up the adjustment dial of the air regulator and then rotate it to adjust the air pressure between 0 3 MPa and 0 5 MPa...

Страница 17: ...e Lubricare 2 can be used with sprays other than those made by Bien Air Dental SA To do this the stand for the spray can and the coupling must be replaced Spray Cans and Cleaning and Lubrication Lines...

Страница 18: ...Check the spray can volume and make sure the protrusion hub in the middle of the spray stand is correctly facing up or down de pending on the spray can as shown in the table below and then set it str...

Страница 19: ...g handpieces Doing so will damage the Lubricare 2 and handpieces Make sure the lock levers are all the way up Otherwise improper connection results in not supplying oil The corresponding Empty Spray C...

Страница 20: ...e height rotate the adjusting screw located in the center of the spray stand with a hex wrench 1 Turn the lock lever down in clockwise direction 2 Check that the lock lever is horizontal and then remo...

Страница 21: ...matched Set the coupling s thick pipe to the right side and in sert the coupling into the connector p 10 5 2 1 Recommended Mode 2 Push the coupling against the connector with one hand and pull down t...

Страница 22: ...is oriented towards the front side of the unit The sheet can be installed either way up Oil Absorbent Pad Oil collection tray is not properly inserted Oil Collection Tray Oil Absorbent Sheet Shorter L...

Страница 23: ...place Magnets Front Door Front Door Line up the magnets with the metal bumps of the sheet stopper Front Door Sheet Stopper Door Oil Absorbent Sheet 1 Insert the front door sheet stopper into the door...

Страница 24: ...d Micromotor Attachment speed increase Chuck Lubrication daily use Chuck maintenance performed before handpiece body lubrica tion Approx 4 sec p 25 7 3 3 Chuck Lubrica tion Handpiece Body Cleaning Lub...

Страница 25: ...ines 1 to 4 are set for Mode 1 Check the handpiece type and change the mode as neces sary p 10 5 2 1 Recommended Mode p 19 6 3 Couplings p 28 7 3 5 Switching Maintenance Mode 7 3 2 Handpiece Body Clea...

Страница 26: ...e Start Stop Switch The Line Selection Lamp for the current cleaning process will blink and the lamps for the reserved lines will light up Press the Start Stop Switch again to stop the process Do not...

Страница 27: ...o not remove the handpiece from the chuck nozzle until you hear the single sustained beep Otherwise it may result in excessive scattering of oil If the nozzle skirt comes off be sure to reinstall it a...

Страница 28: ...d for Line 1 To cancel the line selection press the Line Selection Switch again The Line Selection Lamp will go out 2 Starting Lubrication Face the bur insertion hole towards the main unit If the bur...

Страница 29: ...g the operation This will result in oil leakage and malfunction Do not block the exhaust opening or the exhaust outlet on the bottom of the unit Otherwise the front door will open during the maintenan...

Страница 30: ...ired Line Selection Switch and press the Start Stop Switch for about two seconds until you hear a single sustained beep The Line Selection Lamp and currently set Lubrication Mode Selection Lamp will b...

Страница 31: ...from inside the handpiece 2 Hold down the Line Selection Switch for the line where the handpiece is attached Air Blow Mode starts after approximately two seconds You can release the switch when the p...

Страница 32: ...ound and the Line Selection Lamp for the completed line and the Completion Lamp will light up Wipe excess oil from the handpiece with a soft cloth or gauze after the maintenance 3 Check that the selec...

Страница 33: ...Door Sheet Stopper Order parts from your local dealer or Bien Air Dental SA OFFICE Reinstall all parts after cleaning Do not use disinfectants that contain chlorine based cleaners or benzalkonium chlo...

Страница 34: ...a problem with the air compressor and it may result in poor maintenance Contact your local dealer or Bien Air Dental SA OFFICE p 13 6 1 2 Air Tube Air Regulator Front Door Hold the device firmly with...

Страница 35: ...d with excessive oil absorption Doing so could cause more oil mist to come out from the device and spray into your eyes Dispose of used door oil absorbent sheets according to regulations for medical w...

Страница 36: ...rings 8 3 6 Spray Cans 1 Remove the old O rings with a pair of tweezers or another such tool 2 Set the new O rings according to coupling grooves Replace the can when the corresponding Empty Spray Can...

Страница 37: ...rt the hinge straight until it snaps on p 32 The door seal is partially removed Reinstall the door seal properly N A Process stops before lubri cation Selected Line Selection Lamp and corresponding Em...

Страница 38: ...the lamp still lights up even after performing maintenance again have the device repaired N A Oil seeps out excessively from the connection parts and the handpiece head during normal use of the handpi...

Страница 39: ...on rm that the device is uncontaminated and must then have it disposed of by a healthcare facility or an agent licensed and quali ed to handle standard industrial waste and industrial waste requiring...

Страница 40: ...38 Oil Absorbent Pads Qty 5 Code No 7373465 Oil Absorbent Sheets Qty 5 Code No 7376847 Door Oil Absorbent Sheets Qty 5 Code No 7376839 Nozzle Skirt Qty 1 Code No 7373791 10 3 Consumables...

Страница 41: ...s and appearance are subject to change without notice due to improvements Used to maintain optimum performance and prolong useful life of dental handpieces Delivers oil and air automatically to handpi...

Страница 42: ...ged person people in dental clinic b Language Understanding English or other languages offered in the Instruction for Use Understanding of precautions and warnings c Experience Not relevant 11 2 Inten...

Страница 43: ...imitation Humidity limitation Atmospheric pressure limi tation Prescription Device CAUTION Federal law restricts this device to sale by or on the order of a dentist Valid only for U S A cTUVus certifi...

Страница 44: ...ce CISPR 11 Group 1 Class B This device uses RF energy only for its internal function There fore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipme...

Страница 45: ...line s to earth Signal input output 1 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environ ment Voltage dips short interruptions and voltage variations...

Страница 46: ...ter manufacturer E is the compliance level in V m and d is the recommended separation distance in meters m Field strengths from field RF transmitters as determined by an electromagnetic site survey a...

Страница 47: ...45 None Essential Performance Cable Length No Interface s Max Cable Length Shielding Cable Classification 1 AC Power Cable 2 3 m 3 wire Un shielded AC Power Line...

Страница 48: ......

Страница 49: ......

Страница 50: ......

Страница 51: ......

Страница 52: ...el 1 949 477 6050 Fax 1 949 477 6051 ba usa bienair com Bien Air France S rl 19 21 rue du 8 Mai 1945 94110 Arcueil France Tel 33 0 1 49 08 02 60 Fax 33 0 1 46 64 86 68 ba f bienair com Bien Air Italia...

Отзывы: