6.2 Cleaning
Preparation
1. Disconnect the device from the coup-
ling and remove the bur (
step
1).
Clean under running water
step 1
1. Clean under runing water at <38°C.
Remove dirt / deposits
step 2
1. Carefully remove all traces of dirt or
deposits from the nozzles using the
Bien-Air cleaning wire.
2. Dry the nozzles with compressed air
then dry the rest of the instrument
with a cloth.
CAUTION
Do not immerse in an ultrasonic
cleaner.
Internal cleaning with aquacare
1. Shake the Aquacare before use.
2. Use the Aquacare attachment (see
step 1).
3. Keep the tip of the instrument point-
ing down and protect with an
absorbent cloth; press the Aquacare
button down for 2 seconds (see
step 2).
4. Use the Aquacare attachment (see
step 3).
5. Keep the tip of the instrument point-
ing down and protect with an
absorbent cloth; press the Aquacare
button down for 1 second (see
step 4).