15.1 Enter Endodontics mode
Tap
to enter the endodontics mode. Predefined endodontics
systems are available by default, it is possible to modify and
restore them. It is possible to create and delete User-defined
endodontics brands, systems and files.
15.2 Operative screen description
The operative screen displays the selected operation brand,
system type and file settings:
(1)
Access to brands, systems and file settings
(2)
File name and color designation of file according to ISO
3630-1 standard
(3)
Micromotor rotation mode: FORWARD, AUTO-REVERSE,
AUTO-FORWARD
(4)
Save (to customize settings)
(5)
Footpedal mode: PROGRESSIVE or ON/OFF mode
(6)
Light intensity
(7)
Handpiece ratio
(8)
Maximum micromotor torque in Ncm, indicative percentage of
maximum torque that can be delivered is also displayed
(or according to settings, see chapter “13.7 Settings” on page 45)
(9)
Maximum micromotor speed in rpm
(10)
Figure of the file
(11)
File system name
(12)
File brand
(13)
Back to the previous screen
15.2.1 Micromotor Speed & Torque
Slide to adjust the maximum speed and torque values.
Or tap the icons
to enter a precise value.
Enter the motor speed or torque values and tap
Cancel
or
Done
to
validate.
When the power demand of the micromotor is excessive, the
operative screen displays the overheating symbol
. In this
case the iOptima unit lowers the torque automatically in order to
avoid overheating of the micromotor.
To restore 100% torque, allow the motor to idle or stop for a few
seconds.