5 Operation

5.1 Operating conditions and

procedure

Before operating the device, follow this
procedure:

1. Check that the ambient temperature

ranges b10°C to +35°C
(+50°F to +95°F) and that it is under
standard humidity conditions (re-
lative humidity below 80%, without
condensation).

2. Check that the settings parameters

of the irrigation system are set so
that the irrigation flow supplies the
blade at least 60 ml/min. (Note: If the
internal irrigation is used in com-
bination with a Chiropro console, this
corresponds to the irrigation level: « 
2 drops »).

3. If required by the clinical operation, in-

sert and push the irrigation ring on
the microsaw to set the irrigation
ring on the cylindrical part of the mi-
crosaw (as shown in the

FIG. 1

, the

irrigation ring must be pushed at
least until the end of the front-side ir-
rigation tube of the irrigation ring is
aligned with the last knurled pat-
tern).

4. Connect the irrigation line to the ir-

rigation tube of the microsaw device
and, if the irrigation ring is mounted,
to the 2 irrigation tubes of the ir-
rigation ring.

5. Fix a suitable Bien-Air cutting tool to

the microsaw (which should already
be equipped with the irrigation ring,
if necessary) and orient the irrigation
tube of the microsaw and the ir-
rigation ring so as to ensure a
comfortable grip.

6. Connect the microsaw system (after

having mounted the irrigation ring
and the cutting tool) to a micro-mo-
tor compatible with ISO 3964.

7. Operate without load the microsaw

system for at least 5 seconds and
check that it performs properly.

CAUTION

Check the Microsaw handpiece for dam-
age and loose parts before each use.

WARNING

The use of a coolant while the in-
strument is in operation is mandatory,
failure

to

comply

with

this

re-

commendation

may

result

in

overheating and burns or product fail-
ure. Set the coolant quantity for the
cooling irrigation to at least 60 ml / min.

CAUTION

Depending on the angular position of
the clamping ring, a difference of flow
rate may be noticed for high flow rate.

CAUTION

Before each use, always check that the
physiological solution flows freely and
without leak out of the cutting tool.

CAUTION

Perform a test run during 5 seconds be-
fore each use. If the irrigation of the
surgical site will be realized through the
cutting tool, also activate the irrigation
during the test run and check the ir-
rigation flow.

10

Содержание 1601098-001

Страница 1: ...Rx Only RECIPROCATING MICROSAW ENG INSTRUCTIONS FOR USE REF 2100358 0001 2021 08...

Страница 2: ...RECIPROCATING MICROSAW RECIPROCATING MICROSAW REF 1601098 001 Irrigation ring REF 1502918 001 O rings REF 1308430 001 Optional accessories REF Spraynet REF 1600036 006 Lubrifluid REF 1600064 006 Aquac...

Страница 3: ...2 5 3 2 Connecta new irrigationline 12 5 4 Changing the O rings 13 5 4 1 Missing or damaged O rings 13 5 4 2 Replacing the O rings 13 6 Cleaning and servicing 14 6 1 Maintenance General in formation 1...

Страница 4: ...x code for product information including UDI Unique Device Iden tification CAUTION hazard that could result in light or moderate injury or damage to the device if the safety instructions are not cor r...

Страница 5: ...g to the specific clinical application 2 4 Intended User The microsaw is intended to be used by oral and maxillofacial surgeons and plastic surgeons in an operating room 2 5 Intended medical condition...

Страница 6: ...ove dangerous WARNING Medical personnel using or performing maintenance on medical devices that are contaminated or potentially con taminated must comply with universal precautions in particular the w...

Страница 7: ...ice the dispersion affecting these quantities is below or equal to 5 Technical data Reciprocating microsaw 1601098 001 Irrigation ring 1502918 001 Weight 118 g 12 3 g Overall length 117 mm 34 9 mm 19...

Страница 8: ...the irrigation ring REF1502918 and the irrigation line REF1500984 in combination with motor MX i LED REF1600755 electronic con sole Chiropro L REF1600613 and cutting tool REF1502679 with motor speed 4...

Страница 9: ...table Bien Air cutting tool to the microsaw which should already be equipped with the irrigation ring if necessary and orient the irrigation tube of the microsaw and the ir rigation ring so as to ensu...

Страница 10: ...rd parties WARNING The use of a non original Bien Air cut ting tool will prevent proper irrigation of the surgical site through the orifice of the cutting tool 11 FIG 2 Description of pictogramsused W...

Страница 11: ...has been re moved before reprocessing the device inspect the irrigation ring Make sure the two O rings are in the grooves Cut off an irrigation line portion of 50 60mm about 2 inches with sterile scis...

Страница 12: ...ool such as a dental hook or tweezers to extract the O ring from its groove 2 Carefully pull the O ring out Avoid scratching the metal 3 Insert the new O ring by gently pressing it between your finger...

Страница 13: ...tergents that are pH 8 11 are neither corrosive nor contain chlorine acetone and or al dehydes l Do not submerge in physiological liquid NaCl nor use saline solu tion to keep the device moist until it...

Страница 14: ...icrosaw from the motor 2 Remove the cutting tool see FIG 5 3 Disconnect the irrigation line from the microsaw see FIG 5 4 If present the irrigation ring may be either cleaned separately or left on the...

Страница 15: ...duct at low concentration and at room temperature for a dur ation of 5 10 minutes is recommended for example Neodisher Mediclean Forte at 0 5 for 5 6 minutes 3 Check the freedom of movement of the cla...

Страница 16: ...disinfector re spect the previous program for the Pre cleaning Cleaning Neutralization and Rinsing phases If the local tap water has a pH outside the range of 6 5 8 5 or if it contains more than 100...

Страница 17: ...the spray for about 1 second 4 Clean the excess oil on the exterior with a sterile non woven compress Note If during the test run of the mi crosaw before the clinical treatment an excess of lubricant...

Страница 18: ...he autoclave chamber does not exceed 137 C 278 6 F i e the nominal temperature of the autoclave is set at 134 C 273 2 F 135 C 275 F or 135 5 C 275 9 F taking into account the un certainty of the steri...

Страница 19: ...ckaging 6 7 Servicing Never dismantle the device For all ser vicing or repair operations you are advised to contact your usual supplier or Bien Air Dental SA directly Note The microsaw should be retur...

Страница 20: ...uests will only be taken into consideration if the product is ac companied by a copy of the invoice or delivery note The following information must be clearly indicated purchase date product reference...

Страница 21: ...Bien Air Dental SA L nggasse 60 Case postale 2500 Bienne 6 Switzerland Tel 41 0 32 344 64 64 Fax 41 0 32 344 64 91 dental bienair com Other adresses available at www bienair com...

Отзывы: