5 Maintenance and servicing
5.1 Maintenance – General information
CAUTION
l
Non-sterilizable.
l
Never submerge the hose in disinfectant solutions (the connectors should
never be completely submerged).
l
Do not use an ultrasonic cleaner.
5.2 Cleaning
Clean with a clean cloth moistened with either tap water, sterile demineralized (de-
ionized) water or any appropriate product for dissolving protein and blood
residues.
5.3 Rinsing
Remove the product residues with a clean cloth soaked either in tap water or in
sterile demineralized (deionized) water.
5.4 Drying
Spray the exterior of the hose with Spraynet then remove its excess with a non-
woven cloth. Do not use products containing acetone, chlorine or bleach.
5.5 Packing and storage
Storage conditions
Temperature range:
0°C — +40°C (+32°F — +104°F)
Relative humidity range:
10% — 80%
Air pressure range:
650 hPa — 1060 hPa
Keep away from rain
The device must be stored in a dry and dust free environment.
CAUTION
If the medical device has been stored refrigerated, allow it to warm up to room
temperature prior to its use.