52
ADDITIONAL INFORMATION
Brief fault finding notes and a list of replacement parts are given in this section.
Note:
that this equipment should only be dismantled by properly trained personnel. Removing the
outer cover exposes potentially lethal mains voltages. There are no user serviceable parts within this
equipment.
FAULT FINDING
Should you have any problems with your TC-512 which cannot be easily remedied, you should contact
your supplier and return the unit if necessary. Please include details of the fault observed and remember
to return the unit in its original packing. Bibby Scientific Ltd accept no responsibility for damage to units
which are not properly packed for shipping: if in doubt, contact your supplier, giving the full serial
number of the unit and software version number (shown when the unit is first switched on).
FUSES
If neither the power light nor display on the front panel is lit, one of the two fuses may have blown.
Check that there is no external cause, such as a faulty plug or lead. Check both fuses and replace the
faulty fuse with a new one of the correct value (fuse values are given on the label next to the power
inlet). Note that fuses should only be replaced by a qualified electrician.
The holder for the two fuses is built into the mains input socket. First remove the power cable and then
gently prise the fuse drawer open with a flat-bladed screwdriver or similar tool.
Each fuse can be removed by using the screwdriver as a lever.
Exchange the faulty fuse in the fuse holder for a working fuse of the correct value. Finally, replace the
fuse drawer in the fuse compartment and push the drawer shut.
Fuses which blow repeatedly indicate a serious fault and you should return the unit to your supplier for
repair.
ACCESSORIES
Product no
Description
Quantity
FTUB05TW
0.5ml thin wall microtubes
Pack of 1000 tubes with caps
FTUB02TW
0.2ml thin wall microtubes
Pack of 1000 tubes with caps
FMITUBC
0.2ml thin wall microtubes
250 strips of 8 tubes
FMICAPC
0.2ml microtube caps
250 strips of 8 caps
F96T02
96-well flat top, non-skirted plates
Pack of 25 plates
F96PLHS
96-well half skirted plates
Pack of 25 plates
F96SEAL
96-well sealing mat (silicon)
Pack of 50 mats
FMIP384
384-well skirted plates
Pack of 50 plates
FHSEAL
Heat sealing film
Pack of 100
TC512 Techne Manual 19/12/08 3:30 PM Page 54