Page 5 of 20                                                                                                M6627   Issue 7.2 

 

3. 

SAFETY INFORMATION. 

 

This product has been designed for safe operation when used as detailed in accordance with 
the Manufacture

’s instructions. 

 

NOTE:

 Failure to use this equipment in accordance with the manufacture

’s operating instruction 

may compromise your basic safety protection afforded by the equipment and may invalidate the 
warranty / guarantee. The warranty / guarantee does not cover damaged caused by faulty 
installation or misuse of the equipment. 

 

3.1.  Prevention of Fire and Electric Shock. 

 

 

To prevent a risk of fire or electric shock, 

DO NOT

 open your 

product case without authorisation. Only qualified Service 
personnel should attempt to repair this product   

 

 

 

Replace fuses only with the type as listed in section, Parts and 
Accessories and Technical Specifications. (See fuse type and 
rating) 

 

 

 

Ensure the Mains Power Supply conforms to rating found on the 
data plate located on the back of this Product. 

 

 

 

Never

 Operate this equipment with out connection to earth / 

ground. Ensure the mains supply voltage is correctly earthed / 
grounded in accordance with current area legislation.  

 

3.2.  General Safe Operating Practice. 

 

 

Always follow good laboratory practice when using this  
equipment. Give due recognition to your company’s safety and 
legislative health & safety procedures and all associated 
legislation applicable to your areas of operation. Check 
laboratory procedures for substances being heated and ensure 
all hazards (e.g. explosion, implosion or the release of toxic or 
flammable gases) that might arise have been suitably 
addressed before proceeding.  When heating certain 
substances the liberation of hazardous gases may require the 
use of a fume cupboard or other means of extraction. 

 

 

 

Do not

 position the product so that it is difficult to disconnect 

from the mains supply.  

 

 

 

Do not 

immerse unit in water or fluids.

 

 

 

 

Do not 

spill substances onto this unit. If spillage does occur, 

disconnect unit from mains supply and follow instruction as 
detailed in Section ‘Maintenance’.  

 

 

 

Do not 

cover this unit whilst in use.

 

 

 

 

It is 

not

 recommended to leave any heating apparatus 

unattended during operation. 

 

 

 

Only use Original Equipment manufacture

’s spares and 

accessories. Ref Section 10.  

 

 

 

The equipment is not spark, flame or explosion proof and has 
not been designed for use in hazardous areas in terms of BSEN 
60079-14:1997. Keep flammable, low flash point substances 
away from the apparatus. 

Содержание Electrothermal MH6616

Страница 1: ...Page 1 of 20 M6627 Issue 7 2 SLIDE DRYING BENCH MH6616 X1 INSTRUCTION BOOK...

Страница 2: ...tion of Conformity Page 20 Appendix A Decontamination certificate Page 16 The copyright of this instruction book is the property of Bibby Scientific Limited This instruction book is supplied by Bibby...

Страница 3: ...unting stage Accepting up to 50 slides it has the facility for drying slides in different orientations The temperature control is provided by a built in energy regulator controlling temperature up to...

Страница 4: ...o one s self or other people The Responsible Body and the Operator should read and be familiar with this Instruction book in order preserve the protection afforded by the equipment To prevent injury o...

Страница 5: ...islation 3 2 General Safe Operating Practice Always follow good laboratory practice when using this equipment Give due recognition to your company s safety and legislative health safety procedures and...

Страница 6: ...7 2 Do not operate or handle any part of this product with wet hands Do not touch the heating surface whilst in use Do not lean or stretch over equipment Keep the Mains Plug and Lead set cable away f...

Страница 7: ...n against the diagram For your future reference please record your products Serial and Model numbers Unit Serial Number Unit Model Cat Number Item No Description Qty 1 Mains Cord set Illustration only...

Страница 8: ...mains supply is to be earthed grounded in accordance with current legislation 5 1 3 Ensure only the correct rated mains input fuses are fitted Where applicable ensure the correct mains cable fuse if...

Страница 9: ...rmation For end users outside of the EU consult applicable regulations 6 4 This product should only be dismantled for recycling by an authorised recycling company This product and accessories must be...

Страница 10: ...or 8 Temperature control regulator knob and On Off switch 9 Mains input IEC socket 10 Data plate label 7 2 Place the slides to be dried on the bench surface The slides can be arranged on the bench in...

Страница 11: ...3 x 1 Mains Supply Power 230V AC or 115 100V AC 50 60Hz Mains Power Lead set UK 13A BS1362 3 core earthed ground 2 meters long Moulded IEC plug and Lead set supply cord H05 V V F Replace only with eq...

Страница 12: ...al misuse will reduce the effective life of this product and may cause a hazard Maintenance for the unit should include Periodic electrical safety testing an annual test is recommended as the minimum...

Страница 13: ...it from the power source 9 3 2 Unscrew and remove the screws retaining the top plate Lift the top plate clear 9 3 3 Hinge back the heater mat and insulation pressure plate the insulation should be car...

Страница 14: ...s Before using any cleaning or decontamination method the Responsible Body should check with the manufacturer the proposed method will not damage the equipment Prior to further use the Responsible Bod...

Страница 15: ...Slide Support Rod Qty 4 AZ9139 AZ9139 Rubber Feet Qty 4 AZ9140 AZ9140 Neon Clear AZ9141 AZ9142 Neon Amber AZ9143 AZ9144 Handles Qty 2 AZ9145 AZ9145 Energy Regulator AZ9127 AZ9128 Control Knob AZ9133...

Страница 16: ...leting YES NO as applicable and by providing details in section 2 below A Blood body fluids Pathological specimens YES NO Provide details if YES B Biodegradable material that could become a hazard YES...

Страница 17: ...ain Tel 44 0 1785 812121 Fax 44 0 1785 810405 General enquiries info bibby scientific com Order enquiries sales bibby scientific com Technical support electrothermalhelp bibby scientific com www elect...

Страница 18: ...Page 18 of 20 M6627 Issue 7 2 12 NOTES...

Страница 19: ...Page 19 of 20 M6627 Issue 7 2...

Страница 20: ...fic Limited Beacon Road Stone Staffordshire ST15 0SA Great Britain Tel 44 0 1785 812121 Fax 44 0 1785 810405 General enquiries info bibby scientific com Order enquiries sales bibby scientific com Tech...

Отзывы: