Automatic Handpiece Cleaning & Lubrication System

CareMaster

®

Instructions For Use

Содержание CareMaster

Страница 1: ...Automatic Handpiece Cleaning Lubrication System CareMaster Instructions For Use...

Страница 2: ...Oil Absorbent Sheet Connecting Power Cord Operation Procedure 7 Attaching Handpiece Cleaning and Lubricating Removing Handpiece Door 10 How to remove the Door How to mount the Door Maintenance 11 Rep...

Страница 3: ...horised dealer If the Control Unit smokes or smells of burning turn off the power immediate and disconnect the power plug Contact your dealer Do not drop the Control Unit Place the Control Unit on a f...

Страница 4: ...o perform uniform handpiece maintenance and lubrication Three handpieces can be cleaned and lubricated simultaneously whilst keeping your hands clean CareMaster s E type Connector effectively and effi...

Страница 5: ...To check the Fuse rating remove the Fuse box Refer to 6 Changing Fuses 2 Connecting Air Tube Step 1 Securely insert A side into the back of the unit fig 4 Step 2 Connect CareMaster to the air source T...

Страница 6: ...ressor and Air Filter need to be clean and free from moisture Drain the Air Compressor and Air Filter at least once a week as required Moisture from the tank could mix with the Maintenance Oil and def...

Страница 7: ...Fig 8 equipo Fig 8 CAUTION Be sure to remove the bur from the handpiece before connecting it to each connector 4 Operation Procedures 1 Attaching Handpiece Attaching air turbine handpiece according t...

Страница 8: ...tor until it locks Fig 11 3 Point the bur insertion hole toward the Door and close the Door securely Fig 12 CAUTION Be sure to remove the bur from the Contra Angle Handpiece before attaching it to the...

Страница 9: ...te lubrication and cleaning cannot be performed in the Short or Long mode Speed Increasing Contra Angle Handpiece Speed Reducing Contra Angle Handpiece Other Makes of Air Turbines Contra Angle Handpie...

Страница 10: ...and lubrication cycle has completely finished Pull out the handpiece in axial direction Fig 15 CAUTION Wipe off the Maintenance Oil collected at the bottom of the cleaning chamber and the Door The col...

Страница 11: ...replace with new Mist Filters as shown in Fig 18 Order No HPLP011 CAUTION Replace Mist Filters after approx every 50 cycles Dispose of used Mist Filters as medical waste according to the laws and reg...

Страница 12: ...s medical waste according to the laws and regulations of your country 3 Replacing O ring 1 Push out the worn O ring with your fingers 2 Place new O ring into the empty groove 4 Replacing Air Filter 2...

Страница 13: ...and unplug the Power Cord 2 Squeezing the upper and lower ratchets of the Fuse Box inward at the same time pull out and remove the Fuse Fig 24 3 Replace the Fuse with a new recommended Fuse Refer 3 1...

Страница 14: ...press the start button Does the Control Unit produce an operating sound Not buzzer If the Control Unit produces an operating sound it has no fault Check the Air Tube or the air supply If the Control U...

Страница 15: ...ns Beyes will without charge repair or replace at Beyes sole discretion any defective part s covered in this warranty If the conditions are met the defective item must be returned to an authorized Bey...

Страница 16: ...a Inc 23 595 Middlefield Road Toronto Ontario M1V 3S2 Canada Tel 1 855 603 1888 Fax 1 855 720 1228 Email info beyes ca Web www beyes ca Lotus NL B V Koningin Julianaplein 10 Ie Verd 2595AA The Hague N...