Cleaning and Lubrication
LED Light Cleaning (fig.5)
Clean the LED light using isopropyl alcohol
or other disinfection agents. Never use a
cleaning agent that has an ammonium,
chlorine, or acid base.
External Cleaning
Clean debris from the external surface by
Manual Lubrication
After cleaning the handpiece apply
lubricant
BEFORE
autoclaving
(Fig.6). Use only a high quality
lubricant such as Beyes Lube.
Ensure the correct application
nozzle is connected to the plastic
valve at the top of the spray can.
Firmly insert the application nozzle
into the rear of the handpiece and
activated the spray for 2-3 seconds
until the spray exhausts from the
handpice head. Wipe any excess
lubricant from the handpiece.
Sterilization
The handpiece must be cleaned and sterilized between patient uses.
A wrapped handpiece should be steam sterilized at 132°C for a minimum of 10 minutes
or at 121°C for a minimum of 50 minutes. Do not exceed 135°C.
An unwrapped handpiece should be steam sterilized at 132°C for a minimum of 8
minutes or at 121°C for a minimum of 30 minutes. Do not exceed 135°C.
Note: An unwrapped handpiece needs to be used immediately or sterility is not
maintained.
Do not exceed the sterilizer manufacturer’s guidelines for load capacity.
CAUTION: Use steam sterilization exclusively for complete sterility. Never dry heat
sterilize, ultrasonically clean, or expose handpiece or swivel to chemical disinfection
or cold sterilization by:
• Immersion techniques or Surgical milk
• Surface wipes, Hand soaps, or Detergents
• Surface sprays
• Sterilization bags containing disinfectants
Troubleshooting
Problem: NO LIGHT, REDUCED LIGHT, OR INTERMITTENT LIGHT
Cause: Air Pressure is below the minimum pressure 35 PSI
Correction: Increase the air pressure in your delivery unit.
Problem: YELLOWISH LIGHT
Cause: Debris on LED surfaces
Correction: Clean all exposed optical surfaces. (Refer to Cleaning.)
Limited Warranty
Beyes Dental Canada hereby warrants against defect in material and workmanship
that for a period of 12 months (six months for turbine) from the date of purchase,
this product will perform satisfactorily under normal use and service. No warranty
claim can be inferred here, as wear may occur earlier or later than indicated above
depending on use, frequency of sterilization and frequency of maintenance.
Beyes shall under no circumstances be liable for incidental or consequential
damages or damage due to improper use or maintenance.
This Warranty is subject to the following conditions
Beyes will, without charge repair or replace (at Beyes sole discretion) any defective
part/s covered in this warranty.
If the conditions are met, the defective item must be returned to an authorized
Beyes service center or distributor. Accompanied by the original invoice or
packing list. This warranty is in lieu of all other warranties, expressed or implied.
Use only faultless burs and diamonds. These must be clean to prevent dirt from being
transferred into the chuck.
CAUTION: Never depress the autochuck end
cap button or rest the end cap against any
part of the oral cavity during operation.
Depressing the end cap while a bur is rotating
can activate the chuck mechanism and result
in the end cap button excessive heat and/ or
loss of bur retention.
Bur Insertion (fig.3)
1. Insert bur into autochuck until resistance is
felt.
2. Depress theautochuck end cap button and
continue to feed-in the bur until it clicks and
stops.
3. Release the autochuck end cap button.
(Always pull on the bur after it has been
installed to confirm that it is seated properly).
4. Never depress the autochuck end cap when
the bur is rotating.
Bur Removal (fig.4)
1.Grasp handpiece firmly so your thumb
touches the end cap of the autochuckand push
until a slight ‘click’ is noticed.
2. Pull bur straight out.
3. Release end cap.
fig.4 Bur removal
fig.5 LED light Cleaning
using isopropyl alcohol or other disinfecting agents. Never use a cleaning agent
that has an ammonium, chlorine, or acid base. Dry thoroughly. Never immerse the
handpiece in disinfection baths. Improper care and maintenance may lead to a
premature wear and failure of the handpiece.
Automatic Handpiece Lubrication
After cleaning the handpiece simply connect the handpiece to the correct adaptor
and activate the Lubrication System according to the manufacturer’s system
instructions.
fig.6 Manual Lubrication
Beyes Dental Canada Inc.
9-595 Middlefield Road
Toronto Ontario M1V 3S2
Canada
Product complies with MDL
ISO 13485: 2003 CMDCAS Registered
Spray nozzle
Beyes Lube
File#: M800V1
fig.3 Bur Insertion
push
push
push
pull
Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner
licensed by the law of the states in which he or she practices to use or order the use of this device