Cleaning and Lubrication 

LED Light Cleaning  (fig.5)

Clean  the  LED  light  using  isopropyl  alcohol 

or  other  disinfection  agents.  Never  use  a 

cleaning  agent  that  has  an  ammonium, 

chlorine, or acid base.

External Cleaning

Clean  debris  from  the  external  surface  by 

Manual Lubrication

After cleaning the handpiece apply 

lubricant 

BEFORE 

autoclaving 

(Fig.6).  Use  only  a  high  quality 

lubricant  such  as  Beyes  Lube. 

Ensure  the  correct  application 

nozzle  is  connected  to  the  plastic 

valve  at  the  top  of  the  spray  can. 

Firmly insert the application nozzle 

into the rear of the handpiece and 

activated the spray for 2-3 seconds 

until  the  spray  exhausts  from  the 

handpice  head.  Wipe  any  excess 

lubricant from the handpiece.

Sterilization 

The handpiece must be cleaned and sterilized between patient uses.

A wrapped handpiece should be steam sterilized at 132°C for a minimum of 10 minutes 

or at 121°C for a minimum of 50 minutes.  Do not exceed 135°C.

An  unwrapped  handpiece  should  be  steam  sterilized  at  132°C  for  a  minimum  of  8 

minutes or at 121°C for a minimum of 30 minutes.  Do not exceed 135°C.

Note:  An  unwrapped  handpiece  needs  to  be  used  immediately  or  sterility  is  not 

maintained.

Do not exceed the sterilizer manufacturer’s guidelines for load capacity.

CAUTION:  Use  steam  sterilization  exclusively  for  complete  sterility.    Never  dry  heat 

sterilize, ultrasonically clean, or expose handpiece or swivel to chemical disinfection 

or cold sterilization by:

•  Immersion techniques or Surgical milk

•  Surface wipes, Hand soaps, or Detergents

•  Surface sprays

•  Sterilization bags containing disinfectants

Troubleshooting

Problem: NO LIGHT, REDUCED LIGHT, OR INTERMITTENT LIGHT

Cause:  Air Pressure is below the minimum pressure 35 PSI 

 

Correction: Increase the air pressure in your delivery unit.

Problem: YELLOWISH LIGHT

Cause:  Debris on LED surfaces   

 

Correction: Clean all exposed optical surfaces.  (Refer to Cleaning.)

 

Limited Warranty 

Beyes Dental Canada hereby warrants against defect in material and workmanship 

that for a period of 12 months (six months for turbine) from the date of purchase, 

this product will perform satisfactorily under normal use and service. No warranty 

claim can be inferred here, as wear may occur earlier or later than indicated above 

depending  on  use,  frequency  of  sterilization  and  frequency  of  maintenance.  

Beyes  shall  under  no  circumstances  be  liable  for  incidental  or  consequential 

damages or damage due to improper use or maintenance.

This Warranty is subject to the following conditions 

Beyes will, without charge repair or replace (at Beyes sole discretion) any defective 

part/s covered in this warranty. 

If the conditions are met, the defective item must be returned to an authorized 

Beyes  service  center  or  distributor.  Accompanied  by  the  original  invoice  or 

packing list. This warranty is in lieu of all other warranties, expressed or implied.

Use only faultless burs and diamonds.  These must be clean to prevent dirt from being 

transferred into the chuck.

CAUTION:  Never  depress  the  autochuck  end 

cap  button  or  rest  the  end  cap  against  any 

part  of  the  oral  cavity  during  operation.  

Depressing the end cap while a bur is rotating 

can activate the chuck mechanism and result 

in the end cap button excessive heat and/ or 

loss of bur retention.

Bur Insertion (fig.3)

1.  Insert  bur  into  autochuck  until  resistance  is 

felt.

2.  Depress  theautochuck  end  cap  button  and 

continue  to  feed-in  the  bur  until  it  clicks  and 

stops.

3.  Release  the  autochuck  end  cap  button.  

(Always  pull  on  the  bur  after  it  has  been 

installed to confirm that it is seated properly).

4. Never depress the autochuck end cap when 

the bur is rotating.

Bur Removal   (fig.4)

1.Grasp  handpiece  firmly  so  your  thumb 

touches the end cap of the autochuckand push 

until a slight ‘click’ is noticed.

2. Pull bur straight out.

3. Release end cap.

fig.4 Bur removal

fig.5 LED light Cleaning

using isopropyl alcohol or other disinfecting agents. Never use a cleaning agent 

that has an ammonium, chlorine, or acid base. Dry thoroughly. Never immerse the 

handpiece  in  disinfection  baths.  Improper  care  and  maintenance  may  lead  to  a 

premature wear and failure of the handpiece.

Automatic Handpiece Lubrication 

After cleaning the handpiece simply connect the handpiece to the correct adaptor 

and  activate  the  Lubrication  System  according  to  the  manufacturer’s  system 

instructions.

fig.6 Manual Lubrication

Beyes Dental Canada Inc.

9-595 Middlefield Road                     

Toronto Ontario M1V 3S2             

Canada

Product complies with MDL    

ISO 13485: 2003 CMDCAS Registered

                                                   

Spray nozzle

Beyes Lube

File#: M800V1

fig.3 Bur Insertion

push

push

push

pull

Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner 