Cleaning and Lubrication

 

LED Light Cleaning

  (fig.5)

Clean the LED light using isopropyl alcohol or 
other disinfection agents. Never use a clean

-

ing agent that has an ammonium, chlorine, or 
acid base.

External Cleaning

Clean  debris  from  the  external  surface  by 
using  isopropyl  alcohol  or  other  disinfecting 

Manual Lubrication

After  cleaning  the  handpiece  ap

-

ply  lubricant  BEFORE  autoclaving 
(Fig.6). Use only a high quality lubri

-

cant such as Beyes Lube. Remove 
the handpiece from the tubing and 
Place 2 – 3 drops of BeyesLube in 
the drive air tube of the handpiece. 
While looking at the connector end 
(threaded  end)  of  the  handpiece, 
the  Drive  Air  tube  is  the  smaller  of 
the two large tubes. The larger tube 
is  the  Exhaust  Tube  (no  lubricant 
required).  Insure  the  lubricant  is 

Sterilization 

The handpiece must be cleaned and sterilized between patient uses.
A wrapped handpiece should be steam sterilized at 132°C for a minimum of 10 min

-

utes or at 121°C for a minimum of 50 minutes.  Do not exceed 135°C.
An unwrapped handpiece should be steam sterilized at 132°C for a minimum of 8 min

-

utes or at 121°C for a minimum of 30 minutes.  Do not exceed 135°C.

Note: 

An unwrapped handpiece needs to be used immediately or sterility is not main

-

tained.

Do not exceed the sterilizer manufacturer’s guidelines for load capacity.

CAUTION: 

Use  steam  sterilization  exclusively  for  complete  sterility.    Never  dry  heat 

sterilize, ultrasonically clean, or expose handpiece or swivel to chemical disinfection or 
cold sterilization by:
• 

Immersion techniques or Surgical milk

• 

Surface wipes, Hand soaps, or Detergents

• 

Surface sprays

• 

Sterilization bags containing disinfectants

Troubleshooting

Problem:

 NO LIGHT, REDUCED LIGHT, OR INTERMITTENT LIGHT

Cause: 

Air Pressure is below the minimum pressure 35 PSI 

 

Correction:

 Increase the air pressure in your delivery unit.

Problem: 

YELLOWISH LIGHT

Cause: 

Debris on LED surfaces   

 

Correction: 

Clean all exposed optical surfaces.  (Refer to Cleaning.)

 

Limited Warranty 

Beyes  Dental  Canada  hereby  warrants  against  defect  in  material  and  workman

-

ship  that  for  a  period  of  12  months  (six  months  for  turbine)  from  the  date  of 
purchase,  this  product  will  perform  satisfactorily  under  normal  use  and  service. 
No  warranty  claim  can  be  inferred  here,  as  wear  may  occur  earlier  or  later  than 
indicated  above  depending  on  use,  frequency  of  sterilization  and  frequency  of 
maintenance.  Beyes shall under no circumstances be liable for incidental or con

-

sequential damages or damage due to improper use or maintenance.

This Warranty is subject to the following conditions 

Beyes will, without charge repair or replace (at Beyes sole discretion) any defec

-

tive part/s covered in this warranty. 
If  the  conditions  are  met,  the  defective  item  must  be  returned  to  an  authorized 
Beyes service center or distributor. Accompanied by the original invoice or pack

-

ing list. This warranty is in lieu of all other warranties, expressed or implied.

Use only faultless burs and diamonds.  These must be clean to prevent dirt from being 
transferred into the chuck.

CAUTION: 

Never  depress  the  autochuck 

end  cap  button  or  rest  the  end  cap  against 
any  part  of  the  oral  cavity  during  operation.  
Depressing the end cap while a bur is rotat

-

ing  can  activate  the  chuck  mechanism  and 
result  in  the  end  cap  button  excessive  heat 
and/ or loss of bur retention.

Bur Insertion 

(fig.3)

1.  Insert  bur  into  autochuck  until  resistance  is 
felt.
2.  Depress  theautochuck  end  cap  button  and 
continue  to  feed-in  the  bur  until  it  clicks  and 
stops.
3. Release the autochuck end cap button.  (Al

-

ways pull on the bur after it has been installed 
to confirm that it is seated properly).
4. Never depress the autochuck end cap when 
the bur is rotating.

Bur Removal  

 (fig.4)

1. Grasp handpiece firmly so your thumb touch

-

es the end cap of the autochuckand push until 
a slight ‘click’ is noticed.
2. Pull bur straight out.
3. Release end cap.

fig.3 &4  

Bur Insertion and removal

fig.5 

LED light Cleaning

agents.Never use a cleaning agent that has an ammonium, chlorine, or acid base.
Dry thoroughly. Never immerse the handpiece in disinfection baths.Improper care 
and maintenance may lead to a premature wear and failure of the handpiece

Automatic Handpiece LubricationAfter cleaning the handpiece simply connect the 
handpiece to the correct adaptor and activate the Lubrication System according 
to the manufacturer’s system instructions.

placed in the correct tubeand then run the handpiece for 3 – 5 seconds. Wipe any 
excess lubricant from the handpiece.

fig.6 

Manual Lubrication