18 

Beurer GL43 mg/dL

 Important 

• Do not use third-party control solution. Correct functioning of the measuring device can only be 

tested using the Beurer control solutions ( LEVEL6).

•  Control solution measurements: When using the device, specialist personnel must follow statutory 

guidelines. 

•  Do not apply any blood samples or control solutions to the test strip before inserting it in the moni-

tor.

Performing a functional test using control solution

 Warning 

To obtain correct results, the measuring device, the test strip and the control solution must be the 

same temperature. For the “Functional test using the control solution”, the temperature is to be be-

tween 20 °C and 26 °C.

1

Hold the blood glucose monitor so that the display is facing you.

2

Insert a test strip into the slot on the blood glucose monitor with the contacts first. Make sure 

that the front of the test strip is facing you (see “Test strips” on page 12).

3

The device automatically switches on and briefly shows the initial display. The device is ready 

for measurement as soon as the hand   and the   symbol flash. 

IMPORTANT:

 Control solutions and blood react to temperature influences in different ways. 

It is therefore of vital importance that control solution measurement is always performed in 

control solution mode. If this mode is not used, results may be obtained that are outside the 

target range.

 Note

Press the rocker switch “+” or “–” [2] to change to the control mode. “

” is shown on the 

display. This means that the result value is not stored in the memory, therefore not influencing 

your measured value statistics. Press “+” or “–” [2] again. “

” disappears from the display 

and the value is stored normally in the memory.

4

A clean surface is required to correctly perform a functional test.

Shake the control solution well before use. 

Undo the cap and press two drops next to each other on the 

clean surface without touching them. Use the second drop for 

the measurement.

 Note

Do not apply the drop directly to the test strip to avoid contaminating the remaining control 

solution in the bottle by touching the test strip with the tip of the bottle.

Содержание GL43

Страница 1: ...instructions for use GL43 GB mg dL BLOOD GLUCOSE MONITOR Step by Step Codefree BLOOD GLUCOSE MONITOR tep...

Страница 2: ......

Страница 3: ...surements 15 5 2 Evaluating measured blood glucose values 16 6 Functional check using control solution 17 7 Measured value memory spaces 20 7 1 Displaying individual values 20 7 2 Displaying average b...

Страница 4: ...ood samples either for self testing or in a clinical environment by trained personnel It enables you to measure your blood glucose quickly and easily store the measured values and dis play the average...

Страница 5: ...F H D A 1 Blood glucose monitor B 1 lancing device C 5 test strips D USB cable E AST cap with groove for taking blood samples from alternative parts of the body F 1 button cell battery 3 V CR2032 alr...

Страница 6: ...lled measured blood glucose values from the last 7 14 30 and 90 days Set the time and date Transfer stored measured values to a PC for evaluation The blood glucose monitor also includes the following...

Страница 7: ...display the blood glucose content in mg dL The unit of meas urement mg dL accompanies each blood glucose value Please contact Customer Services if the device does not display mg dL You risk damaging y...

Страница 8: ...pressure shock or hyperosmolar hyperglycaemic non ketotic coma may lead to incorrect measured results A hematocrit value between 20 and 60 has no significant influence on the measurement results An ex...

Страница 9: ...values Notes on handling batteries If your skin or eyes come into contact with battery fluid flush out the affected areas with water and seek medical assistance Choking hazard Small children may swall...

Страница 10: ...nfection of other persons After use dispose of test strips and lancets in a puncture proof container Note The empty completely flat batteries must be disposed of through specially designated collectio...

Страница 11: ...cription of units and accessories 3 1 Blood glucose monitor Front 1 Slot for test strips 2 Rocker switch 3 Display 4 Highlight button 5 On Off button 6 PC port Rear 1 Battery compartment bottom side C...

Страница 12: ...od glucose unit mmol L not functional 9 Blood glucose unit mg dL 10 Hand symbol 11 Test strip symbol 12 Blood droplet symbol 13 Time Note The blood glucose monitor is supplied with the following basic...

Страница 13: ...lass symbol Test strips expire three months after the box is opened Make a note of the expiry date opening date 3 months on the label The shelf life is limited to the expiry date see date next to the...

Страница 14: ...ty if the replace battery symbol appears Replace the battery as soon as possible If appears on the display the battery power level is so low that no more measurements can be taken 4 2 Making and chang...

Страница 15: ...hese instructions describe how to prepare and take a blood sample correctly Note Do not apply blood to the sides of the test strips Do not add blood later if the device does not start measurement Remo...

Страница 16: ...dL The warning is displayed for measured results below 20 mg dL The warning is displayed for measured results above 630 mg dL Warning If you suspect that the blood glucose results are incorrect first...

Страница 17: ...on Very high blood glucose level Above 630 mg dL Take another measurement using a new test strip If the result is identical seek medical at tention immediately 6 Functional check using control solutio...

Страница 18: ...y switches on and briefly shows the initial display The device is ready for measurement as soon as the hand and the symbol flash IMPORTANT Control solutions and blood react to temperature influences i...

Страница 19: ...first drop of control solution was not dis posed of The tip of the bottle was not cleaned correctly The bottle was not shaken well enough Rectify the cause and repeat the test Control solution and or...

Страница 20: ...st 7 14 30 and 90 days Note If you have already saved measured values and you reset the date the average values are calculated as from the new period indicates an empty memory for measured values Pres...

Страница 21: ...ood glucose value and the average value are displayed 2 Press 2 repeatedly to display the average value for 7 14 30 and 90 days 3 You can cancel the process at any time To do so press the On Off butto...

Страница 22: ...ment for blood glucose values and are shown on the display mg dL Mem Press 2 repeatedly to display the average value from the last 14 30 and 90 days be fore meals After displaying the average value fo...

Страница 23: ...n You are in the measurements memory By pressing the highlight button 4 you can switch to the different 7 day average values In this way you can get to the desired average value more quickly For examp...

Страница 24: ...lable to download There you will find all the information required for the data transfer and a de tailed description of the software Note An effective evaluation is only possible if you have correctly...

Страница 25: ...e sure that moisture does not enter the test strip insertion slot Do not spray cleaning agent di rectly on the device Do not submerge the device in water or any other fluids and make sure that no flui...

Страница 26: ...ge Repeat the test using a new test strip as soon as the measuring environment measuring device and test strips have reached room temperature 20 C to 26 C Lt Temperature of the measuring environ ment...

Страница 27: ...ices Problem the test does not start after inserting the test strip into the device and applying blood Cause Solution Insufficient blood or test strip not filled correctly Repeat test using a new test...

Страница 28: ...test Each time device is switched on EMC This device complies with the European standard EN 61326 and is subject to specific precautions with regard to electromagnetic compatibility Please note that p...

Страница 29: ...tion of the control solution The control solution is a red solution with the following D glucose level in percentage shares Ingredients Control solution LEVEL 5 Control solution LEVEL 6 D glucose 0 12...

Страница 30: ...glucose concentrations 100 mg dL 5 55 mmol L Within 5 Within 10 Within 15 250 414 60 4 379 414 91 5 413 414 99 8 Results of the system accuracy for combined glucose concentrations between 24 8 mg dL 1...

Страница 31: ...esults Severely lipaemic blood samples were not tested with the Beurer GL43 blood glucose monitor therefore using the device with these samples is not recommended 3 In the case of severely ill patient...

Страница 32: ...in 200 mg dL 0 03 mmol L 0 7 mg dL 0 04 mmol L 1 54 Heparin 60 mg dL 0 10 mmol L 1 1 mg dL 0 06 mmol L 0 37 Ibuprofen 40 mg dL 1 94 mmol L 3 0 mg dL 0 17 mmol L 0 90 Icodextrin 1094 mg dL 1 68 mmol L...

Страница 33: ...ge due to improper operation Wearing parts Cases of personal negligence on the part of the customer The event that the device has been opened by an unauthorised workshop The warranty does not affect t...

Страница 34: ...34 Beurer GL43 mg dL...

Страница 35: ...Beurer GL43 mg dL 35...

Страница 36: ...GL43_mg dL 0416_GB Subject to errors and changes Beurer GmbH S flinger Stra e 218 89077 Ulm Germany www beurer healthguide com www beurer com...

Отзывы: