ors and changes

Содержание GL42

Страница 1: ...german l engineering Beurer GmbH S flinger Stra e 218 89077 Ulm Germany www beurer com instructions for use GL42 GB mg dL BLOOD GLUCOSE MONITOR step by step Codefree...

Страница 2: ......

Страница 3: ...cose measurement 15 5 1 Evaluating the result 15 5 2 Evaluating measured blood glucose values 16 6 Functional check using control solution 17 7 Measured value memory spaces 20 7 1 Displaying individua...

Страница 4: ...team Getting to know your instrument The GL42 blood glucose monitor is intended for fast and simple blood glucose measurement of fresh capillary blood samples either for self testing or in a clinical...

Страница 5: ...ps D AST cap with groove for taking blood samples from alternative parts of the body E 1 button cell battery 3 V CR2032 already inserted F 10 sterile lancet needles G 1 practical case These instructio...

Страница 6: ...average of the labelled measured blood glucose values from the last 7 14 30 and 90 days Set the time and date The blood glucose monitor also includes the following monitoring functions Warning in the...

Страница 7: ...r must display the blood glucose content in mg dL The unit of measure ment mg dL accompanies each blood glucose value Please contact Customer Services if the device does not display mg dL You risk dam...

Страница 8: ...repare for venous blood samples Do not test any severely ill patients using this device The level of Acetaminophen up to 10 mg dL Ascorbic acid up to 4 mg dL Bilirubin up to 20 mg dL Cholesterol up to...

Страница 9: ...d its accessories from impacts humidity dirt marked temperature fluctua tions and direct sunlight Do not store the device test strips and control solution in your vehicle in the bathroom or in a cooli...

Страница 10: ...event injury and infection of other persons After use dispose of test strips and lancets in a puncture proof container Note Batteries must not be disposed of with household waste As a consumer you are...

Страница 11: ...3 Description of units and accessories 3 1 Blood glucose monitor Front 1 Slot for test strips 2 Rocker switch 3 Display 4 Highlight button 5 On Off button Rear 1 Battery compartment bottom side 1 3 2...

Страница 12: ...od glucose unit mmol L not functional 9 Blood glucose unit mg dL 10 Hand symbol 11 Test strip symbol 12 Blood droplet symbol 13 Time Note The blood glucose monitor is supplied with the following basic...

Страница 13: ...lass symbol Test strips expire three months after the box is opened Make a note of the expiry date opening date 3 months on the label The shelf life is limited to the expiry date see date next to the...

Страница 14: ...ty if the replace battery symbol appears Replace the battery as soon as possible If appears on the display the battery power level is so low that no more measurements can be taken 4 2 Making and chang...

Страница 15: ...ructions for use Taking blood samples These instructions describe how to prepare and take a blood sample correctly Note Do not add blood later if the device does not start measurement Remove the test...

Страница 16: ...dL The warning is displayed for measured results below 20 mg dL The warning is displayed for measured results above 630 mg dL Warning If you suspect that the blood glucose results are incorrect first...

Страница 17: ...on Very high blood glucose level Above 630 mg dL Take another measurement using a new test strip If the result is identical seek medical at tention immediately 6 Functional check using control solutio...

Страница 18: ...y switches on and briefly shows the initial display The device is ready for measurement as soon as the hand and the symbol flash IMPORTANT Control solutions and blood react to temperature influences i...

Страница 19: ...first drop of control solution was not dis posed of The tip of the bottle was not cleaned correctly The bottle was not shaken well enough Rectify the cause and repeat the test Control solution and or...

Страница 20: ...st 7 14 30 and 90 days Note If you have already saved measured values and you reset the date the average values are calculated as from the new period indicates an empty memory for measured values Pres...

Страница 21: ...ood glucose value and the average value are displayed 2 Press 2 repeatedly to display the average value for 7 14 30 and 90 days 3 You can cancel the process at any time To do so press the On Off butto...

Страница 22: ...ment for blood glucose values and are shown on the display mg dL Mem Press 2 repeatedly to display the average value from the last 14 30 and 90 days be fore meals After displaying the average value fo...

Страница 23: ...n You are in the measurements memory By pressing the highlight button 4 you can switch to the different 7 day average values In this way you can get to the desired average value more quickly For examp...

Страница 24: ...soft slightly damp cloth water or a mild cleaning solution Dry the device using a lint free cloth Make sure that moisture does not enter the test strip insertion slot Do not spray cleaning agent di r...

Страница 25: ...suring device and test strips have reached room temperature 20 C to 26 C Lt Temperature of the measuring environ ment blood glucose monitor or test strip below the permitted range Repeat the test usin...

Страница 26: ...ices Problem the test does not start after inserting the test strip into the device and applying blood Cause Solution Insufficient blood or test strip not filled correctly Repeat test using a new test...

Страница 27: ...nd is subject to specific precautions with regard to electromagnetic compatibility Please note that portable and mobile HF communica tion systems may interfere with this unit For more details please c...

Страница 28: ...level in percentage shares Ingredients Control solution LEVEL 5 Control solution LEVEL 6 D glucose 0 12 0 35 Non reactive components 99 88 99 65 Standards The Beurer GL42 monitor corresponds to Europ...

Страница 29: ...mol L Within 15 mg dL Within 0 83 mmol L 116 186 62 4 172 186 92 5 184 186 98 9 Results of the system accuracy for glucose concentrations 100 mg dL 5 55 mmol L Within 5 Within 10 Within 15 250 414 60...

Страница 30: ...er upon purchase Cases of personal negligence on the part of the customer Outside interference The warranty does not affect the customer s statutory liability rights To assert a warranty claim within...

Страница 31: ...Beurer GL42 mg dL 31...

Страница 32: ...751 178 0314_GL42_mg dL_GB Subject to errors and changes...