21
10. Technical specifications
Circumference of the cuff
Approx. 25 to 70 cm
Type
EM 29
Weight
– Device
– Device, cuff and cable
– Approx. 90 g incl. batteries
– Approx. 210 g incl. batteries
Electrode size
Approx. 114 x 59 mm
Parameter (500 ohm load)
Output voltage:
max. 50 Vpp / 5.5 V rms
Output current:
max. 100 mApp / 11 mArms
Output frequency: 2 – 110 Hz
Pulse length
60 – 220 μs per phase
Waveform
Symmetric, biphasic rectangular pulse
Voltage supply
4.5 V (3 x 1.5 V AAA, type LR03)
Operating conditions
0 °C to 40 °C, 20 to 65% relative humidity
Storage
0 °C to 55 °C, 10 to 90% relative humidity
We reserve the right to make technical changes to improve and develop the product. If the device is not used
according to the instructions specified, perfect functionality cannot be guaranteed! This device complies with
European standards EN60601-1 and EN60601-1-2 and is subject to special precautionary measures with regard
to electromagnetic compatibility. Please note that portable and mobile HF communication systems may interfere
with this device. More details can be requested from the stated Customer Service address or found at the end of
the instructions for use.
This device meets the requirements of European Directive 93/42/EC for medical products, as well as those of
the Medizinproduktegesetz (German Medical Devices Act). For this device, a functional test and instruction in
accordance with Section 5 of the Medical Devices Operator Ordinance (MPBetreibV) is not required. It is also not
necessary to carry out safety checks in accordance with Section 6 of the Medical Devices Operator Ordinance
(MPBetreibV).