30
Replacing the battery
Please replace the batteries
when the battery status sym-
bol appears in the display [7].
• Open the battery compartment [4]
by pushing the cover in the area
marked with an arrow and sliding
downwards.
• 3 x 1.5 V batteries, type AAA (LR 03 Micro) should be
inserted for operation. Ensure the polarity is correct. Ob-
serve the graphic in the battery compartment.
• Do not use rechargeable batteries.
6. Operation
General notes on use
Do not switch on the device until the cuff is correctly
positioned. See chapter 5.
The shoulder TENS is intended for individual treatment of
pain in the shoulder.
Switching on
Hold down the On/Off button [1] until a short signal sounds
and the LCD display [6] switches on. When the device is
switched on for the first time, program A is automatically
activated.
Selecting a program
The device provides 4 programs, each with 20 pulse inten-
sities. You can find a detailed description in the following
table:
Prog. Dura-
tion
TENS
mode
Fre-
quency
Pulse
length
Mode
A
30 min.
0-10
min.
Acu-
punc-
ture
4 Hz
220 μs
Continu-
ous
10-20
min.
Burst
2 Hz
110
Hz per
pulse)
220 μs
Interrupt-
ed
20-30
min.
Con-
ven-
tional
110 Hz 60 μs
Continu-
ous
B
25 min. Acu-
punc-
ture
4 Hz
220 μs
Continu-
ous
C
25 min. Burst
2 Hz
110
Hz per
pulse)
220 μs
Interrupt-
ed