26
Deleting measur
ed values
• To clear the memory of the relevant user memory, you
must first select a user memory. Start a request for ei-
ther average values or individual measured values and
press and hold the
SET
button for approx. 3 seconds.
All the values in the current user memory are deleted
after 3 brief beeps are output.
• You have the option to delete individual measured val-
ues by pressing the
SET
button when the measured
values are displayed after the measurement.
• To switch off, press the START/STOP button .
• If you forget to switch off the device, it will switch off
automatically after 3 minutes.
8. Cleaning and storing the device and cuff
• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under wa-
ter, as this can cause liquid to enter and damage the device
and cuff.
• If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries. The cuff line
should not be bent sharply.
9. Error messages / troubleshooting
In case of faults, the
message appears in the display.
Error messages can occur when
•
1
: Systolic pressure is not detected,
•
2
: Diastolic pressure is not detected,
•
3
: The cuff is fastened too tightly or loosely,
•
4
: The pump pressure is higher than 300 mmHg,
•
5
: There is a system or unit error.
In the above cases, you must repeat the measurement. Make sure
that the cuff tube is properly inserted and that you do not move or
talk. Re-insert the batteries if necessary, or else replace them.
10. Specifications
Model no.
BM 40
Measurement
method
Oscillometric, non-invasive blood pressure
measurement on the upper arm
Measurement
range
Cuff pressure 0 – 300 mmHg,
systolic 60 – 280 mmHg,
diastolic 30 – 200 mmHg,
Pulse 30 –180 beats/minute
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
inaccuracy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 119 mm x W 109 mm x H 60 mm
Weight
Approx. 257 g (without batteries)
Cuff size
22 to 35 cm
Permissible oper-
ating conditions
+ 10 °C to + 40 °C, 15-93 % relative air hu-
midity (non-condensing)
Содержание bm 40
Страница 84: ...84 C 4 1 5 B LR6 Beurer 1 1 84 2 85 3 89 4 90 5 91 6 93 7 94 8 95 9 95 10 95 11 97 12 97 13 97...
Страница 85: ...85 2 BF WEEE Waste Electrical and Electronic Equipment 21 PAP Storage Transport Operating CE 93 42 EEC...
Страница 86: ...86 5 5 30...
Страница 87: ...87 3...
Страница 88: ...88 WEEE Waste Electrical and Electronic Equipment...
Страница 90: ...90 4 4 AA 1 5 4 Pb Cd Hg 24 SET 2 60 SET 4 x 1 5V AA LR6...
Страница 91: ...91 SET SET 071 95 5 2 3...
Страница 92: ...92 OK 22 35 162 973 30 42 5...
Страница 93: ...93 190 3 5 6 5...
Страница 95: ...95 SET 3 SET 3 8 9 1 2 3 4 300 5 10 BM 40 0 300 60 280 30 200 30 180...
Страница 97: ...97 11 LXCP12 006060BEH 100 240 50 60 0 5 A 6 600 Beurer 2 12 22 35 162 972 30 42 162 973 071 95 072 05 13...