DataPro 500 Operator Guide

 

Troubleshooting 

 

 

Troubleshooting 

General 

Problem 

Possible cause(s) 

Corrective Action 

Display does not 
turn on 

No AC power 

Sensor cable damaged 

Check power switch 

Unplug 25 pin sensor cable to remove 
a possible short-circuit if the cable is 
damaged 

No laser output 
from the gauge 

Indicator is off and laser is off 

No power to the indicator / fuse 
blown 

Cable not plugged in or damaged 

Indicator is on but laser is off 

Laser interlock is open 

Indicator is on and laser light is on 

Shutter signal is open 

If unit was just turned on, and the 
sensor is hot or cold, it can take a 
few minutes for the sensor to bring 
the laser to the proper temperature 

LaserSpeed sensor temperature 
specification exceeded 

Check and replace fuses as required 

Inspect cables for damage, verify 
connections 

Check laser interlock key switch and 
make sure it is on 

Check rear panel shutter interlock 
connection 

Wait a few minutes for the laser to turn 
on 

Cooling / heating may be required. Run 
compressed air or 28-30 °C water 
through sensor’s coolant ports. To 
avoid condensation, the sensor 
temperature should not be allowed to 
go below dewpoint 

Gauge is rebooting 
/ laser on light on 
sensor cycles on & 
off / comm led on 
sensor is 
perpetually blinking 

Not enough power to the gauge 

Make sure a good quality sensor cable, 
of the proper length, is being used. 
Refer to the LaserSpeed handbook for 
information on cables 

Fuse blow 

Fuse blown 

Replace the fuse with 1.5A 250V Type 
T or Slo fuses.  The fuse holder can 
hold two ¼”x 1” or two 5mm x 20mm 
fuses.   

Disconnect all cables before powering 
the DataPro 500 back on.  Reconnect 
cables one at a time to determine the 
source of the problem. 

Display reads zero 

Improper scale factor 

No laser 

Check indicator scale factor 

Check interlocks and shutter 

Improper count or 
rate on the 
indicator 

Scale factors in sensor and 
indicator do not match 

 

Pulse frequency too high 

Correct either indicator scale factor or 
sensor scale factor to match 

Lower the scale factor in the sensor 
and indicator 

 

Part No. MAN/EN5400-0001 

Page 13 of 14 

                Revision B (Jun 2010) 

Содержание DataPro 500

Страница 1: ...DATAPRO 500 OPERATOR GUIDE daily operation and routine maintenance www betalasermike com Manual Part Number MAN EN5400 0001 English Revision B Copyright Jun 2010...

Страница 2: ...f Conformity 5 General Safety Information 6 Intended Use 6 Laser Safety Precautions 6 Labels and Safety Features 7 Introduction 8 Operation 8 Key Switch 8 Programmable Indicator 9 Fuses 12 Cleaning 12...

Страница 3: ...its use and under conditions of its use EXCLUSIONS This warranty does not cover products which have been modified altered or repaired by any other party than Beta LaserMike or its authorized agents F...

Страница 4: ...on earth point using adequately sized cable or other means of connection Where supplied or specified shielded interconnection cables must be employed with this equipment to ensure compliance with the...

Страница 5: ...ide Declaration of Conformity DataPro 500 Operator Guide Declaration of Conformity Part No MAN EN5400 0001 Page 5 of 14 Revision B Jun 2010 Declaration of Conformity Part No MAN EN5400 0001 Page 5 of...

Страница 6: ...ser beam in the optical sensor is very powerful and can permanently damage eyes not protected by laser safety glasses To avoid exposing yourself to hazardous radiation you must take these precautions...

Страница 7: ...1 1986 For a copy of this document write to Laser Institute of America 13501 Ingenuity Drive Suite 128 Orlando Florida 32826 1 800 345 2737 Labels and Safety Features This section acquaints you with t...

Страница 8: ...ator has a front panel LED that can be configured along with the LaserSpeed to indicate status information Operation Key Switch The DataPro 500 front panel key switch provides laser power and beam shu...

Страница 9: ...Depending on the application the monitor is configured with different settings This section describes the general operation of the indicator and how to change settings In general operators do not con...

Страница 10: ...a complete reference Key Parameters for Users This section only pertains to situations where the indicator is programmed with setpoints The counter should be configured with a lockout code so that onl...

Страница 11: ...Press SEL to move to the 100ths position 5 Press RST key 5 times 6 Press SEL to move to the 10ths position 7 Press RST 4 times 8 Press SEL to move to the ones position 9 Press RST key 3 times 10 Pres...

Страница 12: ...l power lines The fuse holder can hold two x 1 or two 5mm x 20mm fuses If a fuse is blown it should be replaced with a fuse of the same type Cleaning Use a soft cloth moistened with water for cleaning...

Страница 13: ...laser to turn on Cooling heating may be required Run compressed air or 28 30 C water through sensor s coolant ports To avoid condensation the sensor temperature should not be allowed to go below dewpo...

Страница 14: ...a contact person to discuss the instrument failure When returning equipment for service it is important to first obtain a Return Material Authorization RMA number The RMA number is needed for proper...