While the patient is suspended, always keep the patients centered
between the base legs.

Do not attempt to roll the lift or casters over floor obstacles while the
user is in the sling and on board.

Use only Bestcare slings specially designed and tested for Bestcare
patient lifts . Do not use slings manufactured by other companies on
Bestcare sit-to-stand patient lift. Bestcare LLC is not responsible for any
consequence resulting from using non-Bestcare products.

Use only Bestcare manufactured or approved parts and components.
Bestcare LLC is not responsible for any adverse consequence resulting
from using parts and components from other manufacturers.

Service and repair of the patient lift should be performed by Bestcare or
its authorized dealers

ONLY

. Bestcare LLC is not responsible for any

consequence resulting from any unauthorized service or repair.

Never attempt to lift over the maximum capacity of the lift and any
accessories used in conjunction with the lift.  The maximum safe working
load for the lift will be clearly labeled on the lift.

DEFINITIONS

& SYMBOLS

In this manual the user refers to the patient or resident and may be used

interchangeably at different times.   Caregiver refers to the operator or
person who is assisting with the transfer.

Содержание BestStand 400 Series

Страница 1: ...en assembled according to the instructions in this manual All operators should have received or read and understood the instructions for safe and proper operation of the patient lift MODELS BestStand...

Страница 2: ......

Страница 3: ...ur Bestcare Lift verify that all merchandise is complete and free from any shipping damage Refuse delivery if the packaging appears to be badly damaged If the merchandise is received damaged or is mis...

Страница 4: ...esent during any patient lifting The sole function of the patient lift is for transferring a patient from one utility beds bathtubs toilets etc to another The lift should not be used to transport or m...

Страница 5: ...stcare LLC is not responsible for any adverse consequence resulting from using parts and components from other manufacturers Service and repair of the patient lift should be performed by Bestcare or i...

Страница 6: ...to stand patient lifts a solution for patient transfers between bed chair or commode The lift reduces injury risk associated with antiquated manual patient lifting practices Competitively priced and c...

Страница 7: ...riatric Size Kneepad Optional Optional Yes Bariatric Size Footrest Optional Optional Yes Buttock Support Strap Optional Optional Optional Kneepad Calf Support Strap Optional Optional Optional Quick Re...

Страница 8: ...asters No Yes Yes Bestcare is committed to continuous improvements of our products therefore the specs dimensions and features listed above are for guidance only and subject to change without prior no...

Страница 9: ...a part is missing Parts List for E SERIES 1 Base 2 Mast 3 Control Box 4 Pendant 5 Actuator 6 Lifting Boom 7 Kneepad Assembly 8 Footrest 9 Castor with Brake 10 Front Castor Parts List for EL SERIES 1...

Страница 10: ...e nuts 1 Attach boom to top of mast Re insert the bolts into the holes at bottom of the mast and tighten the nuts A Attach the bottom actuator to the bracket on the mast and insert pin Insert keeper r...

Страница 11: ...nob two times B Slide L arm inside bracket with slotted bolt head inside the L arm then tighten to secure Repeat these steps for the other side C With both L arms securely attached to the mast slide t...

Страница 12: ...ttery life Battery capacity indicator on battery pack Soft start and stop for lifting actuator Soft and Hardware over current protection Actuator s Over duty protection with sign on LCD Emergency stop...

Страница 13: ...ould be closed REMOVE AND INSERT BATTERY PACK REMOVE press button A and pull out in the up direction Fig 1 INSERT Insert the battery pack into control box as show in Fig 2 3 While pressing the release...

Страница 14: ...REMOVE AND INSERT BATTERY PACK PLEASE SEE PAGE 13 FOLLOW INSTRUCTIONS TO REMOVE AND INSERT BATTERY PACK FROM CONTROL BOX FOR LEGACY SERIES for PERFORMANCE electronics assembly see page 11 Attach the...

Страница 15: ...rst before using the lift KNOW WHAT YOUR BATTERY IS TELLING YOU RED AMBER GREEN GREEN GREEN ON ON ON ON ON FULLY CHARGED ON ON ON OFF OFF NEED CHARGE ON ON O FF OFF OFF OFF CHARGE IMMEDIATELY DON T US...

Страница 16: ...Press the UP or DOWN button on the hand control once and check if the battery indicator lights are on Position the base of the lift around or under the object Widen the base and engage the caster brak...

Страница 17: ...e patients feet are firmly on the platform and knees are touching the kneepads with kneecaps just above the padding Use the lift to complete the lift and pivot transfer FITTING SUPPORT STRAP DIAGRAM 1...

Страница 18: ...position as long as the patients body clears bed or chair and vice versa you can move the lift and place patient in new position It is not uncommon for the strap to loosen and drop down as lift procee...

Страница 19: ...s have them hold their arms around the chest Press the UP button on the hand control Before the patient s body is lifted from the chair stop and make sure the sling is secured and patient s knees are...

Страница 20: ...raps to hooks in the lower ends of the lifting arm MAINTENANCE INSPECTION The operator of the lift shall inspect the BestStand Lift before each use Check all bolts for tightness Make sure the base can...

Страница 21: ...on the casters for deflection FIRST RECEIVED MONTHLY EVERY 3 MONTHS BASE FOOT PEDAL Apply grease to caster ball bearings if needed Check welding joints for rust and crack CLEANING AS NEEDED ACTUATOR C...

Страница 22: ...g on usage The actuator will either go up or down but not both Replace your hand control as it is at the end of its life cycle Your control box may be malfunctioning There is a grinding sound inside t...

Страница 23: ...tended only to the original purchaser user and is therefore non transferrable Any warranty claims must be submitted through an authorized dealer or distributor that sold the original product with prop...

Страница 24: ...marked with the RA number or the shipment will not be received or any credit issued When a part is returned to Bestcare LLC for warranty inspection a 50 service charge will apply if the part is not de...

Страница 25: ...Bestcare LLC 2920 Pacific Drive Norcross GA 30071 tel 678 679 6690 free 1 877 822 9033 www bestcarellc com 2013 Bestcare LLC...

Отзывы: