20

Maintenance & Inspection

• The Bestcare Lift requires a minimum of service to keep it in proper working order and to 

assure safe operation. It is important that certain basic checks be performed periodically 

by the operator of the lift or by a qualified bio-med or maintenance staff.

• The manufacturer recommends that the following components and operating points 

be scheduled for periodic inspection at intervals not greater than 6 months (≤ 6). The 

manufacturer also recommends visual inspection prior to use for the additional points 

below (V). The periodic inspection interval and the visual inspection points are listed 

below.

• Note – Bestcare assumes that the assembly instructions were performed properly and 

by a qualified individual. Bestcare also assumes that the operator of the lift has been 

properly trained, reviewed all the training material and knows how to operate the lift.

• With proper use and care, the expected lifetime of the lift is 10 years or 20,000 cycles.

• The expected lifetime of the electrical components are 3 years.
• Contact Bestcare for any recycling information

Maintenance

 

&

 

Inspection 

 

The Bestcare Lift requires a minimum of service to keep it in proper working order and to assure safe operation. It 

is important that certain basic checks be performed periodically by the operator of the lift or by a qualified bio-med 

or maintenance staff. 

The manufacturer recommends that the following components and operating points be scheduled for periodic 

inspection at intervals not greater than 6 months (≤ 6). The manufacture

r

 also recommends visual inspection prior 

to use for the additi

onal points below (V). The periodic 

inspection interval and the visual inspection points are 

listed below. 

Note – Bestcare assumes that the assembly instructions were performed properly and by a qualified individual. 

Bestcare also assumes that the operator of the lift has been properly trained, reviewed all the training material and 

knows how to operate the lift. 

 

 

V upon use 

≤ 6 

months 

Boom and Spreader Bar 

 

 

Check connections between boom and spreader bar for proper connection with pin 

& key ring 

(for Patient Lifts)

 

Check connection between boom and mast for improper connection, looseness, or wear 

 

Check spreader bar hook bolts for looseness or damage 

(for Patient Lifts)

 

Key rings are present on 

spreader bar hooks (on patient lifts)

 and on boom hooks 

(on stand assist lifts)

 

Mast 

 

 

Check mast for bending or deflection 

 

Base and Foot Pedal 

 

 

Check all nuts and bolts for looseness 

 

Check all casters and axle bolts for tightness 

 

Check all casters for rotation and locking/unlocking (rear only)

 

 

Apply multi-purpose grease to bearings if needed 

 

Check all weld joints for rust and cracks 

 

Actuator and Control Box 

 

 

Confirm control box is securely affixed to the mast, and battery pack is secured in control 

box 

 

Confirm actuator is secured between the boom and the mast with the pins and key 

rings 

Confirm manual emergency lowering mechanism on actuator operates properly 

 

Confirm red button on control box shuts off all power when pressed 

 

Confirm all cables at the bottom of the control box are properly connected 

 

Sling  

 

 

Inspect the body of the sling and all loops or connectors that could attach to the 

spreader bar for signs of wear, damage, or fraying 

X

 

X

 

Lift Cleaning 

As 

needed 

As 

needed 

 

 

Содержание BESTLIFT PL182

Страница 1: ...F T M o d e l P L 1 8 2 P L 2 2 8 Ensure the product has been assembled according to the instructions in this manual All operators should read and understand the instructions for safe and proper opera...

Страница 2: ..._____________________________________ Date of Purchase _________________________________________ The manufacturing date can be found on the serial number label that is located at the top of the mast T...

Страница 3: ...ional battery pack you will never have to be worried about your lift not being charged and ready for use The PL182 and PL228 are intended to be used with patients who are completely immobile or totall...

Страница 4: ...out its intended Do Not Iron Item may not be smooted or finished with an iron Class II symbol To identify equipment is designed that a connection to electrical earth ground is not required Tumble Dry...

Страница 5: ...ide proper assistance while being lifted i e patients who are comatose spastic agitated or otherwise severely handicapped The patient lift should be used solely for transferring a user patient from on...

Страница 6: ...are exposed that may cause bodily harm or injury Prolonged exposure or physical contact with any part of the equipment WILL NOT cause skin irritation allergic reaction etc Using accessories detachabl...

Страница 7: ...GA 30071 PROPRIETARY AND CONFIDENTIAL THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF BESTCARE LLC ANY REPRODUCTION IN PART OR AS A WHOLE WITHOUT THE WRITTEN PERMISSION OF BESTCARE L...

Страница 8: ...4 43 6 1108 4 43 6 C 802 4 31 6 802 4 31 6 D 582 9 22 9 582 9 22 9 E 963 2 37 9 963 2 37 9 F 852 5 33 6 852 5 33 6 G 547 21 5 547 21 5 H 291 9 11 5 291 9 11 5 I 583 2 23 0 583 2 23 0 J 285 3 11 2 285...

Страница 9: ...ll load Overload Warning by LCD Battery Status by LCD Duty Cycle 10 Duty cycle Max 2 minute ON 18 minute OFF Standard Features 1 6 Point Spreader Bar 2 4 Rear Caster with Brake 3 3 Dual Front Caster 4...

Страница 10: ...Installation 1 Base and Mast 9 Installation 2 Boom 1 3 2 Secure Mast by tightening bolt...

Страница 11: ...10 Installation 3 Actuator Installation 4 Spreader Bar 2 1 It is critical to have the motor part of the actuator to face the same direction as the big arrow as follows...

Страница 12: ...11 Installation 5 Power Control Unit 4 3 2 1 5...

Страница 13: ...ne that serves the needs of the patient while providing the patient with optimal safety Never interfere with the lift unless instructed by the attendant Have a doctor nurse or medical attendant experi...

Страница 14: ...sed improperly Keep long charger cord out of walkway where it poses a tripping hazard Have the charger port on the controller facing away from the wall for easy disconnection When removing the battery...

Страница 15: ...signs below The DC Charger must be connected The emergency stop button must be released The battery will not be charged if the emergency button is pressed in LCD Display Panel Signs When the emergenc...

Страница 16: ...15 LED Indicator Green In Use Blank Standby Lifting Lowering Operating Lift Using Hand Control Operating Lift Using LCD Display Panel...

Страница 17: ...the LCD Display Panel will show the sign on the left either when the emergency stop button is released or when a button on the LCD Display Panel is pressed Lift operation is deactivated Battery needs...

Страница 18: ...lacement with Head Support Polyester with PU Coating M L XL Liko Guldmann Disposable Replacement Nonwoven Disposable M L XL Liko Guldmann Disposable Replacement with Head Support Nonwoven Disposable M...

Страница 19: ...1 Hook up F Point 2 Hook up E Insert F through E 8 Strap Sling A B G H C D E F Point 1 Hook up A C Point 2 Hook up B D Point 3 Hook up E Point 4 Hook up F Point 5 Hook up H Point 6 Hook up G Insert H...

Страница 20: ...t to bed Transfer From Wheelchair to Bed Grasp the sling at each corner of the U shape of the commode aperture The sling should be fitted with the grab handles on the back section facing outward Help...

Страница 21: ...isual inspection prior to use for the additional points below V The periodic inspection interval and the visual inspection points are listed below Note Bestcare assumes that the assembly instructions...

Страница 22: ...previous use Use isopropyl alcohol for any disinfection purpose Cleaning and Disinfecting 21 Cleaning of the lift Cleaning of the sling Solvents or strong liquids or abrasive cleaning products will da...

Страница 23: ...suance of a return authorization will also be refused as the authorization will have expired The final disposition of warranty claims will be subject to our inspection of the returned items Bestcare L...

Страница 24: ...gency Lowering Device is located at the top of the actuator shaft It is intended for use if the actuator or electronics fail to operate while the patient is suspended in midair The device consists of...

Страница 25: ...g on usage The actuator will either go up or down but not both Replace your hand control as it is at the end of its life cycle Your control box may be malfunctioning There is a grinding sound inside t...

Страница 26: ...NOTES 25...

Страница 27: ...NOTES 26...

Страница 28: ...Bestcare LLC 2920 Pacific Drive Norcross GA 30071 Ver 2021 09 www bestcarellc com 2021 Bestcare LLC...

Отзывы: