![Berthold Duo Xpert Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/berthold/duo-xpert/duo-xpert_operating-manual_2739249036.webp)
4 Installation
LB 470 Level
36
56925BA2 Rev. 03, 11/2017
Installation with clamp blocks
j
DANGER
Danger to life from electric shock!
Installation/maintenance may only be carried out if the device has been
de-energised.
Test of absence of harmful voltages when the front side is open.
In case of an electric shock, carry out first aid measures and immediately call an emer-
gency service.
1
2
3
Master EVU
Slave module
Clamp block
4
5
6
Drilling
Guide rail
Fixing screw
Fig. 18
19" subrack with clamp block (Ex: 1x Master, 9xSlave)
NOTICE
The master's EVUs / slave modules must be secured against pulling out by fix-
ing screws (Fig. 18, item 6).
The device must be disconnected from the mains voltage before it is pulled
out.
1.
Set master EVU / slave module (Fig. 18, item 1, item 2) in the guide rails (Fig.
18, item 5).
2.
Carefully slide module into the subrack until the plug connector is inserted
into the clamp block.
3.
Tighten fixing screws (Fig. 18, item 6).
Содержание Duo Xpert
Страница 2: ......
Страница 4: ......
Страница 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Страница 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Страница 183: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 01 11 2017...
Страница 184: ......
Страница 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Страница 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Страница 208: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017 Evaluation unit Auswerteeinheit...
Страница 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...