7 Main Menu Device Setup
LB 478 MPLM
62
56925-8BA2 Rev.00, 07/2019
Interphase Densities (only master EVU)
An interphase denotes the layer with the respective density. In this menu the in-
terphases are defined and a density is assigned to each interphase. The entered
density should be the upper limit value of the density for this interphase. In addi-
tion to the individual interfaces, it is possible to add a total level. The gas density
should be specified as the value for the total level.
1
2
3
Button
< + >
Button
< Edit >
Button
< - >
4
5
6
Table of added interfaces [kg/cm³]
Interface description
Densitiy Value
Fig. 42
Interphase settings
Densities Overview (only master EVU)
Device Setup | Setup | Densities Overview
The control unit reads the detectors' individual local densities in order to build a
density profile of the vessel's content. This density profile might not be perfectly
monotonous (see Fig. 43, item 1), due to statistics, noise and vortexes inside the
vessel. Thus, the measurement algorithm uses mathematical techniques such as a
special averaging / regression to monotonize and slightly smooth the local densities
(see Fig. 43, item 2), before determining the interface positions between the dif-
ferent phases. As a result, interface positions calculated by LB 478 will never be
ambiguous or inconsistent.
1
2
3
Densities
Monotonized densities
Table
Fig. 43
Densities Overview
Содержание Duo Xpert LB 478 MPLM
Страница 2: ......
Страница 4: ......
Страница 12: ...1 About this Operating Manual LB 478 MPLM 12 56925 8BA2 Rev 00 07 2019...
Страница 16: ...2 Safety LB 478 MPLM 16 56925 8BA2 Rev 00 07 2019...
Страница 26: ...4 Installation LB 478 MPLM 26 56925 8BA2 Rev 00 07 2019...
Страница 34: ...5 Electric Installation LB 478 MPLM 34 56925 8BA2 Rev 00 07 2019...
Страница 40: ...6 Operation of the Software LB 478 MPLM 40 56925 8BA2 Rev 00 07 2019...
Страница 74: ...7 Main Menu Device Setup LB 478 MPLM 74 56925 8BA2 Rev 00 07 2019...
Страница 88: ...9 Troubleshooting LB 478 MPLM 88 56925 8BA2 Rev 00 07 2019...
Страница 94: ...10 Maintenance and Repair LB 478 MPLM 94 56925 8BA2 Rev 00 07 2019...
Страница 109: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 03 05 2019...
Страница 119: ...EmulsionSENS Technical Information Technische Information 56925TI1M Rev No 00 07 2019...
Страница 130: ...Technical Information EmulsionSENS Technische Information EmulsionSENS 56925TI1M Rev 00 07 2019...