Chapter 7 Service
73
7.3
Check Measurement with Test Generator
You can use the test generator in code 60 to check the calibration
of the measurement. The test generator simulates the probe and is
enabled by entering a digit larger than 0.
It is advisable to simulate the empty count rate (code 20) as well
as the full count rate (code 21) with the test generator. Correct
switching of the alarm relay can be checked and the reading can
be presented on the display.
If the test generator is turned on:
the device acts as if the count rate were coming from the de-
tector.
the counts from the detector will be ignored.
the error relay is enabled to indicate that the readings do no
longer correspond to the level any more.
Note!
For the limit switch to continue working normally, you have to turn
off the test generator after the test.
To be on the safe side, the test generator will be turned off auto-
matically after 20 minutes.
The test generator is also disabled if the limit switch is locked by
entering the password.
Note!
We recommend documenting all settings in the start-up protocol
on the following pages.
7.4
Plateau Measurement
Plateau measurements are carried out to check the detector in
case of error (only for NaI detector with NaI crystal-multiplier
combination).
The function is started with code 68.
See also the section „Checking the Crystal-Multiplier Combination“
in the hardware manual.
Содержание LB4710-050
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 32: ......
Страница 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Страница 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Страница 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Страница 44: ......
Страница 58: ......
Страница 98: ......
Страница 104: ......
Страница 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Страница 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Страница 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Страница 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Страница 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Страница 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Страница 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Страница 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Страница 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Страница 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Страница 120: ......
Страница 141: ...Mini Switch LB 471 141 Notes...
Страница 142: ...Mini Switch LB 471 142 Notes...
Страница 145: ......
Страница 146: ...Limit Switch Mini Switch LB 471 Software Manual User s Guide detect and identify Process Control...
Страница 147: ......
Страница 149: ......
Страница 151: ......
Страница 161: ......
Страница 195: ......
Страница 215: ......
Страница 219: ......
Страница 223: ...Mini Switch LB 471 78 Notes...
Страница 224: ...Mini Switch LB 471 79 Notes...
Страница 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...