Chapter 15 Troubleshooting and correcting faults
Problem
Cause of problem / action to be taken
Ikus: the graph
Possible causes
display stops
moving, parameters
cannot be adjusted
• faulty communications between control
computer and laptop
• batteries not supplying enough current
• the electronics (main and backup control
computers) have failed
What to do?
Switch the laptop off and then back on again, wait for the
start-up procedure to be completed, then start the monitor
program.
IMPORTANT: The Ikus continues running with the set
parameters.
The graphs remain frozen
The Ikus is operating in emergency pulse mode, see
section 15.3: Emergency pulse mode, page 200.
Notify Berlin Heart.
Restart the Ikus after consulting the service department
staff: See section 15.2: Restarting Ikus, page 198.
Acoustic and visual
Possible causes
alarm from the Ikus,
message
Error: no
data from Master
• simultaneous malfunction of both control
computers
or
Error: no
• power supply malfunction
reaction from
What to do?
Master
Assess the condition of the patient and the hemodynamic
values.
Notify Berlin Heart immediately.
Pump stands still -
Possible causes
no further pump
function; acoustic
• complete failure of the Ikus
alarm is still audible
• batteries are completely empty or serious fault in
the batteries
What to do?
Immediately connect the Ikus to the mains. Until that
supply the patient with the manual pump. See also
section 6.4: Switching between mains and battery
operation, page 79 and 1section 15.5: Driving blood
pump(s) with the manual pump, page 204.
Notify Hotline immediately.
Tab. 15-1
Possible problems
1000721x09 Revision 8
194
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...