Chapter 15 Troubleshooting and correcting faults
3.
If necessary log into the monitor program by entering user ID and password,
confirming the password with
<Enter>
.
4.
In the monitor program, select the option
Pause left
respectively
Drive pause
and press
<Enter>
to confirm. Respond to the prompt in the dialog window by
pressing the
<X>
key or the
<1>
key. The right blood pump will stop.
LVAD:
Pause left
The view
Pump size and single-step mode
is displayed.
BVAD:
Drive pause
The view
Select operating mode
is displayed.
5.
As soon as the pump(s) has/have stopped, clamp off the cannulae beneath
the pump to be replaced and slide the cannulae off the pump. If it is necessary
to clamp any other part of the cannula that is not covered with velour, cover
the part of the cannula that will be clamped with a gauze sponge.
6.
Check cannulae for visible deposits. If necessary, remove these deposits care
fully.
7.
Remove the driving tube of the left pump to be replaced from the connector. To
do so, take hold of the release sleeve and pull this out of the connector.
Connect new left blood pump to the Ikus
INSTRUCTION
1.
Fill the free ends of the cannulae with sterile saline solution. Make sure that all
air has been removed. Connect the prepared replacement pump to the cannu
lae.
2.
Plug the new driving tube into the freed connector. The plug snaps into place
clearly audible.
3.
Check that the plug is securely connected. To do so, grip the plug body above
the release sleeve and pull on it. Do not pull from the release sleeve, and
never from the tube!
4.
Release the tube clamps from the cannulae.
Starting the Ikus
INSTRUCTION
1.
Move the cursor to the field
step left
.
2.
Confirm
Step left
with
<Enter>
to trigger a single step.
3.
If any air bubbles are visible remove them via the de-airing needle. When all
air has been completely removed from the left pump: remove the de-airing
needle.
4.
Move cursor to the
OK
field and press
<Enter>
to confirm. The driving unit
starts up again using the defined parameters.
5.
Check whether the pump is filling correctly and, if necessary, adjust the
parameters.
6.
Secure all connections with cable ties. See section 9.11: Securing the connec
tions, page 130.
15.2 Restarting Ikus
WARNING
Do not switch off and restart the
Ikus
unless the service consultant
requests to do so (e. g. in emergency operating mode).
1000721x09 Revision 8
198
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...