Chapter 7 Commissioning the Ikus and setting parameters
7.2.8 De-airing the blood pumps in single-step mode
NOTICE
Each de-airing step (
Step left
/
Step right
) carries out half a pump cycle
(systole or diastole), the 1
st
step being a diastole. Normally, several de-
airing steps are required for each pump. In single-step mode, the
pumps will operate using the pressures shown in the parameter table.
It will not be possible to switch to the standard view unless at least 1
de-airing step has been completed for each connected pump.
INSTRUCTION
1.
Bring the patient into the Trendelenburg position.
2.
Move the cursor to the field marked
Step left
.
3.
Lift the pump. The de-airing nipple is the highest point.
4.
To trigger a single step, press the
<Enter>
key. If necessary, use the de-airing
needle to vent the air from the pump (see section 8.5: De-airing the blood
pump, page 112). After consulting the surgeon: If necessary, press
<Enter>
repeatedly to trigger further single steps until all air has been removed from
the pump(s). If the blood pump is not filling sufficiently, ensure there is suffi
cient preload and if necessary, increase the diastolic pressure.
5.
In biventricular mode: Move the cursor to the field
Step right
. Repeat the pro
cedure for the 2
nd
pump.
Transition to continuous pumping mode
is not yet possible because
OK
field is
still inactive.
Fig. 7-8
Single-step mode
7.2.9 Starting the blood pump (changing to standard view)
WARNING
Do not start the pump(s) until all air has been removed.
Once the de-airing needle has been removed it cannot be re-inserted.
Only remove the de-airing needle after all air has been removed from
the blood pump, the blood pump is running and the parameters have
been adjusted (see section 7.2.10: Checking the parameters when the
pump is started and adjusting them, page 103 and section 9.10:
Removing the de-airing needle, page 130).
INSTRUCTION
1.
Move cursor to the
OK
field and press
<Enter>
to confirm. The system now
starts with the parameter values visible in the parameter table.
1000721x09 Revision 8
102
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...