Chapter 7 Commissioning the Ikus and setting parameters
7.2.10 Checking the parameters when the pump is started and adjusting
them
WARNING
In order to avoid air being sucked into the blood pump through the
cannula anastomosis, adjust the parameters gradually. If air does enter
the system, disconnect the driving tubes from the Ikus and de-air the
system using the de-airing needle.
Continuously monitor all settings.
Once the de-airing needle has been removed it cannot be re-inserted.
NOTICE
If the pump is not filling adequately at this stage, increase the preload
by adding volume from the CPB circuit. After adding volume, adjust the
parameters on the laptop of the
Ikus
as described in the following table.
INSTRUCTION
1.
Observe the left blood pump. Is the pump ejecting completely? If not: increase
the left driving pressure if necessary.
2.
Observe the right blood pump. Is the pump ejecting completely? If not:
increase the right driving pressure if necessary.
Observe
Action / measure
Right pump
Is the pump filling properly? (see
below)
If not: check the filling pressure (central
venous pressure; CVP)
CVP too low: substitute volume
CVP too high: increase suction
pressure
If no improvement occurs: check the
position of the cannulae via
echographic monitoring!
Left pump
Is the pump ejecting properly?
If not: check mean arterial pressure
(Guideline value: 70mmHg)
Compare left and right pump.
If yes: increase suction pressure on left
Is left pump filling considerably worse
side
than right pump?
If no improvement occurs: check the
position of the cannulae via
echographic monitoring!
Tab. 7-4
Pump filling criteria
1000721x09 Revision 8
103
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...