LEAKSHOOTER
®
LKS1000 is equipped with a special MAX
RMS value function which thanks to the target, shows you if
you are close to the leak (simple round target) or in front of
the leak (round cross in its center).
The method consists of scanning the scene, conscientiously.
From far to close location. Do not forget to decrease GAIN
slowly to avoid saturation.
Start with the maximum of sensitivity (GAIN=106 dB,
settled when the device is switched ON).
Sweep the scene for example from left to right and from
top to bottom, with the aim of going in front of the leak and
thus activating the special MAX value, which will remain
frozen a few seconds (bold vertical line in the bargraph).
If you still have the white target with cross, everywhere,
Leak &
ultrasound
it is because you are in a saturated zone. Look on the
ground with
LEAKSHOOTER
®
, slightly decrease the GAIN to
be in the green part of the bargraph and scan again in such
a way as to find a direction in which to go.
Thus, go back to the area to find this special MAX value
(align RMS real time and MAX). You will see a cross in the
center of the dynamic target. You are in front of the leak.