15
CHROMOPHARE® D 650plus, D 530plus, D 500plus, D 650, D 530, D 500 Service Manual (E)
4 By pressing and holding down Key 4 (DARK) the illuminance is continuously reduced down to the minimum
value. Repeated short presses decrease the illuminance in steps.
5 If the indicator light shows red, there is a fault in the control unit electronics
6 Only with the EndoLite® option. By pressing Key 6 EndoLite
®
the surgical area illumination is switched off and
the surround lighting is switched on
7 Only with the EndoLite® option. By pressing Key 7 (OP light) the EndoLite
®
surround lighting is switched off
and the surgical area lighting is switched on again.
8 Only with the LFN model. By pressing and holding down Key 8 (field size large) the light field is continuously
enlarged up to the maximum value. Repeated short presses enlarge the light field in steps.
9 Only with the LFN model. By pressing and holding down Key 9 (field size small) the light field is continuously
reduced down to the minimum value. Repeated short presses reduce the light field in steps.
Booster activation
1. Press key 4 (Bright) until LED flashes up in the key (max. brightness).
2. The light booster will be activated by pressing simultaneously the key 1 (ON) and 3 (Bright). Luminous power
rises from 110 kLux to 130 kLux.
Note!
The light booster is deactivated by switching off the light with key 2 (OFF).