13
EN
Dimension
153*41*44 mm
Weight
approx 84 g (with batteries)
Batteries
DC3V (2ň batterieAAA)
Battery life
approx 1 year / 6000 readings
Ingress protection rating
IP22
Troubleshooting
Error
Problem
Solution
Er1
Measurement taken befo-
re thermometer is ready
Take temperature when the symbol
or appears on the display
Er2
The room temperature
is not within the range
between 10°C and 40°C
(50°F ~ 104°F)
Place the thermometer in a room for
at least 30 minutes with a temperature
between 10°C and 40°C (50°F ~ 104°F)
Er3
The thermometer is not
correctly placed.
Refer to “Instructions for use” paragraph
for correct use and proceed with a new
reading
Er4
The thermometer detects
a rapid change in the
room temperature
Allow the thermometer to rest in a room
for at least 30 minutes at room tempe-
rature between 10°C and 40°C (50°F ~
104°F) before taking the temperature
Er5
The thermometer is not
functioning properly
Remove batteries, wait 1 minute and
place batteries back in the battery
case. If the error message still appears,
contact distributor
Hi
In forehead mode:
measured temperature
is higher than 43.0°C
(109.4°F)
In object/surface mode:
measured temperature is
higher than 100°C (212°F)
Refer to “Istructions for use” paragraph
for correct use and proceed with a new
reading
Lo
In forehead mode:
measured temperature
is lower than 34.0°C
(93.2°F)
In object/surface mode:
measured temperature is
lower than 0°C (32°F)
Refer to “Istructions for use” paragraph
for correct use and proceed with a new
reading
Low battery
Replace the batteries