In the time that elapses between an impulse and another, the software will show a black screen where 4
graphics appear.
The two numbers which appear in the upper right and left, are the movement counters that determine when
the examination ends. The examination ends after completing 30 moves in the range between 3000 and 6000
° / s
2
angular acceleration for both the right and the left side. However, the doctor may decide to discontinue
the examination when he sees fit.
In the two graphs at the sides, under the movement counter, a histogram appears that shows in detail the
performed movements and tells the doctor which must still be done.
The histograms in red and in blue indicate the movements that are valid for the purposes of the test.
Histograms in gray indicate movements not valid for the purposes of the test, but still useful for patient
monitoring by the physician. Next to each blue, red, gray histogram there is a small green line indicating
the number of positive responses. Highlighted in yellow is the bin of acceleration in which fell the last
performed impulse.