BENNING IT 130
short instructions
- 17 -
2.4 Low-impedance resistance (R
LOW
)/ continuity test
1. Set function
2. Set sub-function and limit.
R LOW
R LOW
Ω
Ω
Ω
Ω
low-impedance resistance (testing
current > 200 mA) or
CONTINUITY
(testing
current < 8.5 mA)
Maximum limiting value
[without (---), 0.1
Ω
÷ 20.0
Ω
]
Buzzer
[ON/OFF] only for CONTINUITY
function
3. Connection diagrams
Connection of the three-wire test cable and of the optional measuring line BENNING TA 5 (044039)
Connection of 3-wire test lead and "Commander" probe tip
4.a R LOW
Ω
Ω
Ω
Ω
low impedance
Press the key
.
4.b CONTINUITY:
Press the key
to begin performing a
continuous measurement.
Press the key
TEST
again to stop measurement.
5. The measuring result can be stored by means of the "MEM" key.
R
...... R LOW
Ω
R+
.... Result at positive polarity
R-
..... Result at negative polarity
R
...... CONTINUITY resistance