Control solution is used to check the performance of the kit.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label or single foil pack for exact range.
Perform a Control Test:
Insert an unused strip
→
Make sure the code number is correct (Ketone)
→
Press
“S” (3 secs)
→
Enter control solution mode (shown CS on screen)
→
Shake the
control solution well
→
Discard first three drops
→
Put one drop onto a clean
surface
→
Touch the control solution with strip sample inlet
→
Test result display
after count down
→
Compare the result with the control range on test strip vial or
single foil pack
Note:
• Do not reuse the test strip.
• Marked open date on new open control solution.
• Do not use the meter if the control test is out of range.
• If the control test keeps result wrongly, please contact an authorized distributor.
• Control solutions are not included. Please contact an authorized distributor for purchasing.
Chapter 5
Control Solution Test
EA13120901
2020-01
EA13120901
2020-01
β-Ketone Test Strip
BK-SK1
0.1 - 8 mmol/L
1048
Control Range LV1
Control Range LV1
0.1-8 mmol/L
Control Range
Code
Model
K1303190
2020/12
BK-SK1
1048
LV1
0.1-8 mmol/L