After the visual inspection, the following tests are carried out for the "Hosp. bed test CI" example:
Please note that only DUTs with a power consumption of up to 3700 VA may be supplied
with power via the UNIMET® 300ST test socket. If DUTs with higher power consumptions
are connected, the thermomagnetic circuit-breaker integrated into the UNIMET® 300ST
power switch will trip. Once the DUT has been removed, the test system can be switched
on after a few seconds.
** ATTENTION! **
Mains voltage at the
socket outlet
Next:
LOAD CURRENT
Value : 25 mA
Limit: --------
- - - Measuring - - -
MAINS VOLTAGE
Value: 230 V
Limit: --------
- - - Measuring - - -
POWER CONSUMPTION
Value: 6 VA
Limit: --------
- - - Measuring - - -
4. Confirm message
The leakage current test is carried out
Load current
< 5 mA ?
no
yes
** WARNING **
Load current <= 5 mA
DUT switched on?
Next:
If applicable:
Switch on device and
then press ENTER key
R-PE
Value: 0.004
Ω
Limit: 0.300
Ω
- - - Measuring - - -
R-ISO CI
Value: > 300.00 M
Ω
Limit: 2.00 M
Ω
- - - Measuring - - -