H5000 Series MultiTherm Shakers

 

 

5

Chapter 1: Introduction 

 

The MultiTherm shaker is an ideal instrument for intensive mixing of samples under 

temperature regulated conditions. 

Mixing and temperature control modes can be used both simultaneously and 

independently i.e, the device can work as shaker and/or as an incubator. The base 

instrument can be used with a variety of different blocks. The instrument is applicable 

for DNA analysis, extraction of lipids and other cell components, DNA library creation, 

PCR amplification, pre-denaturation in electrophoresis, serum solidification etc.

 

 

Features 

of this product as follows

1. 

Because it can be used with various optional mixing blocks, the instrument can 

adapt to different tubes and plates to accommodate many experimental needs. 

Customized blocks are available to suit your specific demands. 

2. 

LCD display   

3. 

Simultaneous display of set and actual time, temperature and speed   

4. 

Over-heat protection system ensures safety & reliability   

5. 

Temperature can be calibrated to meet the user's needs   

6. 

End-of-run signal alerts the user of cycle completion 

 

 

 

Chapter 2: Specifications 

 

1.  Normal operating conditions 

Ambient temperature: 5

30

C  /  Relative humidity: 

70% 

Note: Do not attempt to operate the instrument at temperatures or humidity levels           

outside of this range. 

 

 

Содержание H5000-H

Страница 1: ...MultiTherm Shaker Operations Manual Item H5000 H and H5000 HC...

Страница 2: ...instrument properly please read this manual carefully before operation Please keep it available for future use Prior to the initial operation Please check the Instrument and the accessories against th...

Страница 3: ...elines Use this product indoors Do not operate in a damp or wet environment Do not attempt to disassemble or repair the Instrument by yourself Disassembly will void the warranty If there is a malfunct...

Страница 4: ...environments Keep the instrument and the air vents at least 4 inches from walls other instruments or any obstructions Shut the power off when you finish your work Unplug the machine if it will not be...

Страница 5: ...8 1 Physical Description 8 2 Keyboard and Display panel 9 3 Key Functions 10 Chapter 4 Operation Guide 11 1 Single Temperature Speed and Time setting 11 2 Sequence Link Setting 12 3 Disabling the Temp...

Страница 6: ...oresis serum solidification etc Features of this product as follows 1 Because it can be used with various optional mixing blocks the instrument can adapt to different tubes and plates to accommodate m...

Страница 7: ...C Cooling time N A Block Configuration A BLOCK 96 well Plate or 96 x 0 2ml B BLOCK 0 5ml x 54 C BLOCK 1 5ml x 35 D BLOCK 2 0ml x 35 E BLOCK 15ml x 12 F BLOCK 12mm x 24 G BLOCK 50ml x 6 H BLOCK 0 5ml x...

Страница 8: ...to 100 C Cooling time 30min from room temp to 20 C below room temp Block Configuration A BLOCK 96 well Plate or 96 x 0 2ml B BLOCK 0 5ml x 54 C BLOCK 1 5ml x 35 D BLOCK 2 0ml x 35 E BLOCK 15ml x 12 F...

Страница 9: ...cal description location of controls function of individual keys as well as some preparations before the initial power on You should be familiar with this chapter before the Thermo Shaker is first ope...

Страница 10: ...Shakers 9 2 Keyboard and Display panel Display Panel P1 100 0 1200 10 30 P1 100 0 1200 15 00 Current temperature Current speed Remaining time Selected temperature Selected speed Selected time Current...

Страница 11: ...and hold the or keys Sequence Link Key Press the Seq Link key to run consecutive programs Choose from P1 to P2 P1 to P3 P1 to P4 or P1 to P5 to run up to five consecutive programs The Sequence link fu...

Страница 12: ...rature The initial preset mixing speed and time are also displayed at 1000rpm and 10 00 respectively c Five programs are available for storage By pressing the or Prog keys choose the desired program P...

Страница 13: ...To adjust the ending program press the or Prog Keys Once the ending program has been selected press the Seq Link key to finalize c To begin the first program press the start stop key The first progra...

Страница 14: ...ture calibration The temperature of the Instrument has been factory adjusted and calibrated prior to being shipped But if there is deviation between the actual temperature and the displayed temperatur...

Страница 15: ...low calibration point H5000 H Calibration points 40 C 100 C H5000 HC Calibration points 10 C 40 C 100 C c After 20 mintues read the temperature of the Thermometer for ex 9 8 C Using the or Temp keys a...

Страница 16: ...temperature calibration of the instrument has now been completed and the displayed temperature will accurately represent the temperature of the heating block NOTE During the temperature calibration pr...

Страница 17: ...rench 3mm to remove the four screws that hold the block to the heating board b Close the lid and lift the block up and away from the base unit c Place the desired block softly on the heating platform...

Страница 18: ...atures are significantly different Temperature sensor failure Call for service 3 OPEN in the temperature display with a short beep Temperature sensor failure or the environmental temperature is below...

Страница 19: ...Notes PH 908 769 5555 EM info benchmarkscientific com...