BSH6000 Operation manual

                                                             

  Chapter 4 Operation Guide 

 7 

 

3.

 Temperature Calibration

 

 

The temperature of the instrument has been calibrated prior to shipment from the 
factory.  However, in the event that an adjustemnt to the calibration is required, the 
user can adjust the calibration with either a thermometer/thermocouple. 
 
Caution: the instrument uses double temperatures adjustment to ensure accuracy over 

a wide range of temperature. It is linearly adjusted at both 40° and 100°C.     

 
 
3.1. Calibration with a thermometer,    Block A: 

 

a) Power on the machine and ensure that the temperature on the display is less than 

35°C. 

 
b) Inject olefin oil into one of the cone-shaped wells, and then put a 

thermometer/thermocouple into this well (make sure that the precision of the 
thermometer should be within 0.1

 and the temperature ball should be absolutely 

immerged into the cone-shaped well). See Fig a.     

 

 

 

 

           

 

 

 

 

 

 

 
 

c) Press “Mode” key without releasing, then press 

 key   

at the same time and hold for 3 seconds, it will enter into   
the temperature calibration interface, see the right.   
At this point, “        ” displayed, which indicates entering  
into temperature calibration program. The temperature   
displayed is current temperature and begins to rise to   

 

 

Fig a 

Thermom

Olefin 

il

cone-shape

Содержание BSH6000

Страница 1: ...Operation Manual BSH6000 Iso Block Dry Bath Incubator...

Страница 2: ...DATA ERROR BOOKMARK NOT DEFINED CHAPTER 3 OVERVIEW 3 1 STRUCTURE OVERVIEW 3 2 KEYPADS 4 3 DISPLAY 4 CHAPTER 4 OPERATION 5 1 TEMPERATURE AND TIME SETTING 5 2 ADVANCED TIMER SETTING ERROR BOOKMARK NOT D...

Страница 3: ...lification of DNA the pre denaturation of the electrophoresis blood serum coagulation and a variety of laboratory procedures Features Large digital display of time and temperature Extremely accurate t...

Страница 4: ...humidity 70 Power supply AC115V or 230V 50 60Hz 2 Technical Data Item No BSH6000 Temperature range Ambient Room Temp 5 C 105 C Timer Max 99hr 59min Temperature accuracy 0 1 C Temperature uniformity 0...

Страница 5: ...er focuses on the introductions of the structure keypads and key functions of the instrument as well as preparatory work prior to initial operation 1 Structure overview 2 Power supply Fuse On Off butt...

Страница 6: ...m Temp Time h m Block A Block A Block B Block B Block A Mode Block A Block B Time Start Stop Temp Time Mode Temp Time Time Start Stop TM 3 Display Mode Start Start Stop key After the temperature and t...

Страница 7: ...y are actual settings real time temperature and time remaining The example shows an actual temperature of 28 5 C and 35 minutes remaining c To adjust the set temperature press or keys until the desire...

Страница 8: ...even once time has expired and oVEr is showing on the time display This setting can be adjusted so that the heater will shut down once time has expired To adjust the advanced timer setting pres the M...

Страница 9: ...A a Power on the machine and ensure that the temperature on the display is less than 35 C b Inject olefin oil into one of the cone shaped wells and then put a thermometer thermocouple into this well m...

Страница 10: ...nfirm the input value d Then the instruments will the automatically begin to heat to 100 Once 100C is reched and held as constact the decimal begins to flicker After 20 minutes the user can read the a...

Страница 11: ...se On Off button broken Change button Others Contact the supplier 2 The actual and displayed temperatures are quite different Broken sensor or loose contact of the block Contact the supplier 3 OPEn di...

Страница 12: ...ual Annex 10 Annex Wiring Diagram for reference only KEY a nd Display PT1 Power HEAT1 HEAT2 PT2 MAIN BOARD FAN2 FAN1 Key andDisplay L PE N HEAT PT100 B lock1 HEAT PT100 B lock2 Fan Fan Switch PORT1 J3...

Страница 13: ...Memo...

Отзывы: