background image

 
 

page 2 

Using and Setting the Timer 

The dry bath has a built-in digital timer function that 

sounds a “beeping” alarm and turns off the heat 

when the set time has been reached.  

 

                                   

 

Press the 

Mode 

button to illuminate the 

Timer

 lamp and place the dry bath into the timer set mode.  

Use the 

Up

 and 

Down

 arrow buttons to set the desired time in one minute increments on the 

display. Once time has been set, press and hold the 

STAR/STOP

 button to start the timer. The dry 

bath will 

make a “beep” sound and automatically begin the timer. The display will revert to showing 

the well temperature and the red 

Timer

 lamp will be extinguished.  When the display shows the well 

temperature, pressing the 

Mode

 button will cause the display to show how much time remains on 

the timer in minutes. 
 
When time has expired, the dry bath will briefly sound the 

“beeping” alarm, turn off the heater and 

turn off all lamps. The display will show a set of chara

cters that represent the word “OVER” with 

only a partial R shown.  At this point the user can press the 

START/STOP

 button to start the timed 

cycle over again.  The user can also use the 

MODE

 button to enter the timer mode and change the 

time setting. 
 
It is best to allow the dry bath to first reach the desired temperature before setting and using the 
timer function. ALWAYS SET THE TIMER TO 0000 WHEN USING THE DRY BATH WITHOUT 
THE TIMER FUNCTION AS THIS WILL PREVENT SHUT OFF OF THE HEATER FUNCTION IF 
THE TIMER IS ACCIDENTALLY STARTED. 

 
User Calibration Function

 

The dry baths are calibrated at the factory.  They also have a user calibration capability if needed 
for field calibration.  Please contact the factory to access this feature. 
 

ERR Code on the display 

If the dry bath temperature sensor is shorted or senses an ambient temperature below 0C, the 
display will show the error code ERR. 
 

Warranty: 

The dry bath comes with a 1-year warranty. 

 

 

website:   

www.BenchmarkScientific.com

         

e mail:  

[email protected]

 

   

 

  

Tel:

 

908-769-5555

   

             

Fax: 908-222-1864

 

 

 

 

Rev. 7-10 

 

 

 

 

 

  
    
 
 
 
 
 
 
 
 
                                                                                                                    