Benchmark Scientific Iso-Block BSH6000 Скачать руководство пользователя страница 1

Operation Manual 

 

 
 

BSH6000 

Iso-Block   

Dry Bath Incubator 

 

DOMINIQUE DUTSCHER SAS

All manuals and user guides at all-guides.com

all-guides.com

Содержание Iso-Block BSH6000

Страница 1: ...Operation Manual BSH6000 Iso Block Dry Bath Incubator D O M I N I Q U E D U T S C H E R S A S All manuals and user guides at all guides com a l l g u i d e s c o m...

Страница 2: ...OVERVIEW 3 STRUCTURE OVERVIEW 3 KEYPADS 4 DISPLAY 4 CHAPTER 4 OPERATION 5 1 TEMPERATURE AND TIME SETTING 5 2 ADVANCED TIMER SETTING ERROR BOOKMARK NOT DEFINED TEMPERATURE CALIBRATION 7 CHAPTER 5 FAIL...

Страница 3: ...he electrophoresis blood serum coagulation and a variety of laboratory procedures Features Large digital display of time and temperature Extremely accurate temperature control up to 105 C Built in ove...

Страница 4: ...60Hz Technical Data Item No BSH6000 Temperature range Ambient Room Temp 5 C 105 C Timer Max 99hr 59min Temperature accuracy 0 1 C Temperature uniformity 0 2 C Heating time 20 105 C 20min Block Materia...

Страница 5: ...ructure keypads and key functions of the instrument as well as preparatory work prior to initial operation Structure overview Power supply Fuse On Off button B block Chamber B Block Display Transparen...

Страница 6: ...Start Start Stop key After the temperature and time setting is selected press this key to start the timer This key can also be pressed to stop the timer Down key Used in both time and temperature mode...

Страница 7: ...ng The example shows an actual temperature of 28 5 C and 35 minutes remaining c To adjust the set temperature press or keys until the desired temperature Ex 55 5 is showing on the display After 5 seco...

Страница 8: ...be adjusted so that the heater will shut down once t i meh a se x p i r e d T oa d j u s t t h ea d v a n c e dt i me rs e t t i n gp r e st h e Mo d e k e ya n d h o l di t f o r 1 0s e c o n d s A f...

Страница 9: ...s less than 35 C b Inject olefin oil into one of the cone shaped wells and then put a thermometer thermocouple into this well make sure that the precision of the thermometer should be within 0 1 and t...

Страница 10: ...heat to 100 Once 100C is reched and held as constact the decimal begins to flicker After 20 minutes the user can read the actual value from thermometer N o t e s P l e a s er e a dt h ea c t u a l v a...

Страница 11: ...atures are quite different Broken sensor or loose contact of the block Contact the supplier 3 O P E n d i s p l a y e d a l a r mi n g b e e p Sensor disconnect Contact the supplier S H O r d i s p l...

Страница 12: ...m for reference only PT1 Power HEAT1 HEAT2 PT2 MAIN BOARD FAN2 FAN1 Key andDisplay L PE N HEAT PT100 B lock1 HEAT PT100 B lock2 Fan Fan Switch J3 J12 J1 J8 J6 J10 J11 D O M I N I Q U E D U T S C H E R...

Страница 13: ...Memo D O M I N I Q U E D U T S C H E R S A S All manuals and user guides at all guides com...