BSH5000 Operation manual             

                                         

Chapter 1 Introduction 

 

 

─ 

1

 

 

 

Chapter 1 Introduction 

 

The dry bath incubator is controlled by sophisticated software, which 

can be widely used for the preservation and reaction of samples, the 

amplification of DNA, the pre-denaturation of the electrophoresis, blood 

serum coagulation and a variety of laboratory procedures. 

 

 

Features: 

 

Large digital display of time and temperature   

 

Extremely accurate temperature control up to 105°C 

 

Built-in over-temperature protection 

 

In-Lab calibration 

 

Plastic lid for enhanced precision and eliminating waste of energy 

 

Quick-Flip blocks available for a variety of tube sizes   

 

 

Содержание BSH5000

Страница 1: ...Operation Manual BSH5000 myBlock Dry Bath Incubator...

Страница 2: ...BOOKMARK NOT DEFINED CHAPTER 3 OVERVIEW 3 1 STRUCTURE OVERVIEW 3 2 KEYPADS 4 3 DISPLAY 4 CHAPTER 4 OPERATION 5 1 TEMPERATURE AND TIME SETTING ERROR BOOKMARK NOT DEFINED 2 ADVANCED TIME SETTING ERROR...

Страница 3: ...mplification of DNA the pre denaturation of the electrophoresis blood serum coagulation and a variety of laboratory procedures Features Large digital display of time and temperature Extremely accurate...

Страница 4: ...35 C Relative humidity 70 Power supply AC115V or 230V 50 60Hz 2 Technical Data Item No BSH5000 E Temperature range Ambient Room Temp 5 C 105 C Timer Max 99h59min Temperature accuracy 0 5 C Temperature...

Страница 5: ...ew This chapter focuses on the introductions of the structure keypads and key functions of the instrument as well as preparatory work prior to initial operation 1 Structure overview Housing Power inle...

Страница 6: ...key After the temperature and time setting is selected press this key to start the timer This key can also be pressed to stop the timer Down key Used in both time and temperature mode to decrease the...

Страница 7: ...shown on the display are actual settings real time temperature time remaining The example shows an actual temp of 28 5 C and 35 min remaining c To adjust the set temperature use the and keys to selec...

Страница 8: ...isplay will show OP 1 use the or key to select OP 2 The setting has now been changed and can be stored by pressing the Start Stop key 3 Temperature Calibration The temperature of the instrument has be...

Страница 9: ...then press key at the same time and hold for 3 seconds it will enter into the temperature calibration interface see the right At this point displayed which indicates entering into temperature calibra...

Страница 10: ...ressing or Then press Start Stop to confirm the input value Note Pressing Set and simultaneously during the temperature calibration indicates exiting the temperature calibration program The adjustment...

Страница 11: ...displayed temperature to 39 6 by pressing or Then press to confirm the input value d Then the instrument will heat to 100 automatically Once 100 has been reached and held constant the decimal point wi...

Страница 12: ...e drybath or directly into a sample tube b Press and hold Stop then simultaneously press and hold for 3 seconds the external indicator light goes on to indicate that it has entered into the temperatur...

Страница 13: ...se On Off button broken Change button Others Contact the supplier 2 The actual and displayed temperatures are quite different Broken sensor or loose contact of the block Contact the supplier 3 OPEn di...

Страница 14: ...al Annex 12 Annex Wiring Diagram for reference only L PE N Switch USB PT1000 CONN BOARD J3 J2 J1 AC 220V J4 Heat POWER BOARD J3 Hea tCtl J2 AC9V J2PT100 J4PT1000 USB 1 US B J1 AC 9V X1 He atCtl MAIN B...