APPENDIX: Guidance and manufacturer’s declaration of conformity
(Table according to EN 60601-1-2)
Electromagnetic Emission
Guidance and manufacturer’s declaration – electromagnetic emissions
The BEMER Veterinary Line products for horses are intended for use in the electromagnetic
environment specified below. The customer or the user of the BEMER Veterinary Line device
should ensure that it is used in such an environment.
Emissi st
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The BEMER Veterinary Line device uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The BEMER Veterinary Line product are suitable for use in all establishments,
including domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Electromagnetic Immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The BEMER Veterinary Line products for horses are intended for use in the electromagnetic
environment specified below. The customer or the user of the BEMER Veterinary Line device
should ensure that it is used in such an environment.
Immunity test
standard
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
–
Electrical fast
transient / burst
IEC 61000-4-4
2 kV for power
supply lines 1 kV for
input/output lines
2 kV for power supply
lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
1 kV line to line
2 kV line to earth
1 kV line to line
2 kV line to earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
<5 % U
T
(0.5 cycle)
40 % U
T
(5 cycles)
70 % U
T
(25 cycles)
<5 % U
T
for 5 s
<5 % U
T
(0.5 cycle)
40 % U
T
(5 cycles)
70 % U
T
(25 cycles)
<5 % U
T
for 5 s
Mains power quality should be that of a
typical commercial or hospital environment.
Note: U
T
is the a.c. mains voltage prior to application of the test level.