1) Remove the following items from chair packaging
2) Plug chair into 120VAC outlet.
3) Turn on the main power switch located on the left side of the pump cover. A green lamp will illuminate.
4) Raise the seat by foot switch.
5) Attach the armrest to side frame as shown in Fig 4-2-1.
It is recommended to lightly lubricate armrest shaft with
silicone lube).
6) Attach the seat plate to base plate, using 4 screws.
See Fig. 4-2-2.
Backrest Frame Assembly
•
Metal Seat
•
Plastic Base Cover
•
Small carton box containing Armrests, headrest mechanism and bag of hardware.
•
To avoid the risk of electrick shock, this equipment must only be connected to a supply mains with
protective earth. Grounding reliability can only archieved when the equipment is connected to an
equivalent receptacle marked HOSPITAL ONLY or HOSPITAL GRADE.
Operate the Main Power Switch by hand only. Turn off the main switch after daily operation.
Do not touch a switch on the foot control during the main power switch is being turned on.
CAUTION
CAUTION
NOTE
Fig 4-2-1
Fig 4-2-2
4-2 Installation
Armrest
Side Frame
Armrest Spring
Collar
Flat washer
M6 x 22mm
Armrest Screw
Seat Plate
M6 x 15 Panhead Screws