background image

  

• Motor Position Fault: 

 

 

 

   LED 1, 2 alternating at 0.25 sec. rate.

 

 

    › If  device is unable to stop at the home position  

 

 

      then fault is indicated by beeping for 10 sec.

Battery percentage indications in running state:

 

 

• Battery Charged > 70%: 

 

 

   LED 1, 2, 3 ON.

 

 

• 30% < Battery Charge ≤ 70%: 

 

 

   LED 1, 2 ON.

 

 

• 15% < Battery Charge ≤30%:  

   

 

   LED 1 ON.

 

 

• 1% < Battery Charge ≤ 15%:  

 

 

   LED 1 flash on/off 1 sec. rate.

 

 

• If the battery charge is <1% and the user attempts to  

 

 

   power on the device, LED1 will flash at 0.5 second rate  

 

 

   for 10 seconds and return to off mode.

Battery Charge Indicator in Charging state:

 

 

• Battery Charge > 90%: 

 

 

   LED 1, 2, 3 ON.

 

 

• 70% < Battery Charge ≤ 90%:  

 

 

   LED 1, 2 ON, LED 3 repeat on/off 1 sec.

 

 

• 30% < Battery Charge ≤ 70%: 

 

 

   LED 1 ON, LED 2 repeat on/off 1 sec.

 

 

• Battery Charge ≤ 30%:  

 

 

   LED 1 repeat on/off 1 sec.   

70% < 

Battery 

Charge

3

2

1

30% < 

Battery 

Charge

3

2

1

3

2

1

3

2

1

3

2

1

3

2

1

3

2

1

9    

SkinPen

®

 Precision Rev.V1.0

Over 

Temperature 

Fault

Motor 

Position 

Fault

Clear 

Faults

3

2

1

3

2

1

3

2

1

Содержание SkinPen Precision 100

Страница 1: ...Engineered Designed in the USA Made in the USA Inductive Charging SMART Technology Patented Reciprocating Mechanism User Manual SkinPen Model 100 SkinPen Charger Base Model 101...

Страница 2: ...ectrical Safety Warnings 4 7 Intended Use 5 8 Procedure Instructions Post Procedure Instructions Post Procedure Care 7 9 Cleaning of SkinPen Precision Charger Base 7 10 Storage 7 11 Disposal 8 12 Warr...

Страница 3: ...and a BioSheath Each component and accessory will be explained to understand how SkinPen Precision works SkinPen Precision Key Components A Power Indicator B Power On Off Button C Charge Level Indicat...

Страница 4: ...sing without drugs and or a biologic to protect against abrasion and friction during the microneedling procedure RESCUE Calming Complex Key ingredients assist in reducing the risk of prolonged inflamm...

Страница 5: ...ons Actinic solar keratosis active acne allergies to stainless steel collagen vascular diseases or cardiac abnormalities diabetes eczema psoriasis and other chronic conditions on other areas of the bo...

Страница 6: ...ions warnings and precautions Consult other resources ie IFU for additional information regarding the application of microneedling therapy How to install uninstall disposable SkinPen cartridge Ensure...

Страница 7: ...and up to 2 5mm on the body Lower the setting of the cartridge to 0 25 0 5mm to perform the procedure around the orbital rim Decrease the needle length by adjusting according to the tick marks on the...

Страница 8: ...ONS POST PROCEDURE INSTRUCTIONS POST PROCEDURE CARE For Procedure Post Procedure instructions and Post Procedure Care refer to SkinPen IFU Rev V1 0 9 CLEANING OF SKINPEN PRECISION AND CHARGER BASE Ens...

Страница 9: ...Medical LLC Customer Service at 1 888 372 3982 for warranty inquiries Warranty does not cover Defects due to negligence alteration modification or installation by anyone other than factory authorized...

Страница 10: ...battery charge is 1 and the user attempts to power on the device LED1 will flash at 0 5 second rate for 10 seconds and return to off mode Battery Charge Indicator in Charging state Battery Charge 90...

Страница 11: ...h 15 ENVIRONMENTAL CONDITIONS Operating conditions Temperature 15 30 C Relative humidity 30 75 relative humidity non condensing Transportation conditions Temperature 20 60 C Relative humidity 10 98 re...

Страница 12: ...15223 2012 IEC 60601 1 6 IEC 60529 IEC 62304 ISO 10993 1 ISTA 2A IEC 62133 UN 38 3 This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device...