143
No Purified
Water
Failure to reach operating level with
purified water during the filling process. If,
in the filling process with purified water
(Y13), Operating Level 2 of the level
sensor (P1) is not reached within 7
minutes, the controller terminates the
current program with error display Cause
and remedy see Error 141.
K13
Supply and shut off valves
Check with facilities maintenance for
any water outages. Verify all water supply valves are turned on.
Water valves
Water valves not opening. Activate all water valves in
the service mode and verify activation/flow.
Check voltage to coil & Check 24Vac power supply & Check fuses
F1, F2, F9, F10 & Check wiring, connectors.
Strainers
Clogged water supply strainers. Clean.
Hoses
Check hoses for any kinks or obstructions. Melted rubber
oozes out of SS braids. Hoses lying against steam hoses. Replace
hose(s).
146
No Purified
Water from
Tank
Failure to reach operating level with pre-
heated purified water (M9) from the tank
during the filling process. If Operating
Level 2 is not reached within 3 minutes
while the filling process with pre-heated
purified water (Y14) is under way, the
controller terminates the current program
with error display. Threshold monitoring:
Threshold adjustment for pre-heated DI
water supply.
K14
Supply and shut off valves
Check with facilities maintenance for
any water outages. Verify all water supply valves are turned on.
Water valves
Water valves not opening. Activate all water valves in
the service mode and verify activation/flow.
Check voltage to coil & Check 24Vac power supply & Check fuses
F1, F2, F9, F10 & Check wiring, connectors.
Strainers
Clogged water supply strainers. Clean.
Hoses
Check hoses for any kinks or obstructions. Melted rubber
oozes out of SS braids. Hoses lying against steam hoses. Replace
hose(s).
Содержание WD 290
Страница 1: ...Washer Disinfector Service Training Seminar Belimed US Charleston SC...
Страница 3: ...Automatic vertical glass door Door opens down Use the same rack as the 230 WD 250...
Страница 4: ...WD 290 15 tray washer disinfector with automatic glass sliding doors...
Страница 5: ...Accessories instrument racks...
Страница 8: ...4 key factors for cleaning efficacy Chemicals...
Страница 19: ...TOSI testing...
Страница 27: ...Dynamic fill advantages...
Страница 28: ...WD 290 high voltage section Power Supply Circuit breakers Contactors...
Страница 73: ...Drying system Separation of air and water systems...
Страница 75: ...Drying system components 2 Check valve Box Elements Over temp reset P18 Pressure switch T2 Sensor Exhaust Pot...
Страница 82: ...Service mode...
Страница 83: ...Analysis...
Страница 84: ...Dosing...
Страница 85: ...Controls...
Страница 86: ...Parameter 1 Analysis 2 Dosing 3 Controls 4 Parameter 5 Configuration 6 Limits...
Страница 87: ...Program Structure...
Страница 88: ...Configuration...
Страница 89: ...Limits...
Страница 90: ...WD 250 magnet template...
Страница 93: ...WD 290 controls Status General information incl time and date error messages transport unit status etc...
Страница 96: ...Process status display...
Страница 97: ...Menu...
Страница 98: ...Log on...
Страница 99: ...Password entry...
Страница 100: ...Data entry example...
Страница 101: ...Machine information...
Страница 102: ...Error messages...
Страница 103: ...Settings...
Страница 104: ...Measurements...
Страница 105: ...System configuration...
Страница 106: ...Parameters...
Страница 107: ...Function medium...
Страница 108: ...Door position...
Страница 109: ...Manual function...
Страница 110: ...Magnet positions...
Страница 125: ...PM check sheet...
Страница 126: ...Wash are disassembly...
Страница 127: ...Lower wash arm docking port...
Страница 128: ...Upper wash arm...
Страница 129: ...Daily screen cleaning...
Страница 130: ...Wash Arm...
Страница 131: ...Water supply strainers...
Страница 132: ...Air HEPA pre filter...
Страница 133: ...Drying system...
Страница 134: ...Air water check valve...
Страница 135: ...Air water check valve...
Страница 136: ...Light bulb...