16
5.5
Calibration and calibration mode setting
The electronic scale makes mass measurements using gravity (g). Different geographical
regions and differences in altitude correspond to different values of gravity acceleration (g).
Therefore, to obtain accurate measurements, the scale must be adapted to the place of use
and to the environmental conditions. This adjustment is made via the calibration function.
Press the menu key
to access the main menu.
Use the arrow keys
to move up and
down in the menu and go to section
5
– Calibration of the scale.
Then confirm the selection with the key
.
Before performing the calibration procedure,
check that the plate is empty.
You can calibrate the scale using the default
weight. Or through automatic weight recognition.
Use the arrows
to switch from one
mode to another.
Then confirm the selection with the key
.
Default weight method:
By selecting the default calibration mode when
the calibration command is given, the required
weight will be the factory default one whose value
will vary according to the scale model.
After confirming, the calibration procedure will be
activated.
Load the required calibration weight or press the
key
to cancel the calibration procedure.
Wait for the weight to be acquired.
If the operation is successful, the weighing
screen will be displayed with the value of the
calibrated weight.
Then remove the weight from the plate.