background image

13

5.8 Calibration of the rate display using
      internal references:   SEt

This function enables the zero and span of the rate
display to be adjusted without the need for an
accurate external current source, or for
disconnection from the 4/20mA loop.  The
accuracy of this method depends upon the
accuracy of the internal references which should
be regularly calibrated as described in section
5.9.2

To calibrate the rate display select 'SEt' from the
main menu and press P.  The instrument will
display 'ZErO', pressing P again will reveal the
current display at 4mA.  For flow applications this
will usually be zero.  Each digit of the rate display
can be changed by pressing the Up or Down
buttons.  When the first digit is correct pressing P
will transfer control to the next digit.  When the
least significant digit has been adjusted, press E to
return to the 'ZErO' prompt which completes the
adjustment.

To adjust the rate display at 20mA, press the Up
button which will cause the instrument to display
'SPAn'.  Pressing P  again will reveal the existing
full scale rate display.   Each digit can be changed
by pressing the Up or Down buttons. When the first
digit is correct, pressing P will transfer control to
the next digit. The most significant digit can only be
adjusted by rolling-over the previous digit. When all
the digits have been adjusted press E to return to
the 'SPAn' prompt, followed by E to return to the
main menu.

5.9  Conditioning sub-menu:   Cond

This sub-menu allows the mains (line) frequency at
which the instrument has maximum ac rejection to
be selected and the two internal references to be
calibrated. To minimise the possibility of accidental
adjustment these functions are contained in a
protected sub-menu.  To gain access to this sub-
menu select 'Cond' from the main menu and press
P which will cause the instrument to display '0000'.
Using the Up, and Down buttons and the P button
to move to the next digit, enter the password
'SurE'.  'S' is entered as '5'.  Pressing E will then
give access to the sub-menu.

5.9.1  AC rejection:   FrE

WARNING

If the mains filter frequency is changed, the
rate display and the internal references (if
used) must be recalibrated.

To provide maximum low frequency rejection the
internal digital filter may be set to operate at 50 or
60Hz to correspond with the local mains (line)
frequency.   To change the frequency select 'FrE'
from the 'Cond' sub-menu and press P which will
reveal the current setting.  The setting can be
changed by pressing the Up or Down  buttons
followed by the E button to return to the sub-menu.

5.9.2  Calibration of internal references:
          rEF

The BA558C contains two references representing
4 and 20mA.  These are used in the 'SEt' function
that enables the rate display to be calibrated
without an external current calibrator.  They are
also used in the 'C--P' function when the P push-
button is programmed to display the input current
in the display mode.   If either of these two
functions are used, the internal references should
be routinely calibrated against an external 4/20mA
calibrator connected to the input terminals of the
instrument.   See section 7.4

To calibrate the internal references select 'rEF'
from the sub-menu and press P  which will result in
a '.004A' prompt being displayed. Adjust the
external current calibrator to 4.000mA  and again
press P.  The instrument will display 'Ent' while the
4mA reference is being updated and will then
return to the '.004A' prompt.

To re-calibrate the 20mA internal reference, press
the  Up button which will cause the instrument to
display '.020A'. Adjust the external current
calibrator to 20.000mA  and again press P.  The
BA385C will display 'Ent' while the 20mA reference
is being updated and will then return to the '.020A'
prompt.  This completes calibration of the internal
references, two operations of the E button will
return the instrument to the 'Cond' prompt in the
main menu.

5.10  Resetting grand total:   CLr. Gtot

The grand total is a separate sixteen digit counter
which duplicates the total display, but is not zeroed
when the total display is reset to zero.  The grand
total may be viewed in the display mode in two
eight digit sections as described in section 2.1 of
this manual.

Содержание BA558C

Страница 1: ...Issue 3 1st October 2010 BA558C 4 20mA loop powered panel mounting rate totaliser issue 3 Label Rate display Annunciators for optional alarms Rotating flow indicator Total display...

Страница 2: ...factor 5 14 Percentage clip off 5 15 Local reset of total display 5 16 Security code 5 17 Over and under range 6 Calibration Examples 6 1 Using an external current source 6 2 Using the internal refer...

Страница 3: ...applied to the instrument initialisation is performed After a short delay the following display sequence occurs All segments of the display are activated for about 3 seconds Decimal points on large d...

Страница 4: ...ators plus a flow indicator as shown on page 1 Rate Shows the flow rate on the smaller display five digit display Total Shows total flow since the display instrument was last reset on the larger eight...

Страница 5: ...instrument provides IP65 protection and a gasket seals the joint between the instrument enclosure and the panel The instrument may be installed in any panel providing the environmental limits shown i...

Страница 6: ...IP65 seal between the instrument enclosure and the panel c Connect the panel wiring to the rear terminal block s as shown in Fig 5 To simplify installation the terminals are removable so that panel w...

Страница 7: ...isplay may be calibrated to show the flow rate in engineering units Both the zero display at 4mA and the span display at 20mA are adjustable although for flow application the display at 4mA is usually...

Страница 8: ...econds while the new information is stored in permanent memory All new BA558C rate totalisers are supplied calibrated as requested at the time of ordering If calibration information is not supplied th...

Страница 9: ...liamps or the input as a percentage when the P push button is operated in the display mode Select PC for of span and 4 20 to display current See section 5 12 Display Description of function SCALE t To...

Страница 10: ...cimal point clip off to adjust value Press or between 0 and 10 or and pressing Define Security Enter by Code to move to next digit Local reset or Fig 7B Programme Structure CLr Gtot E E E E E E E E P...

Страница 11: ...or and Similarly set the required full scale display accurate 20mA input current scale display using an Similarly set the required full to move to next digit pressing or and set required zero display...

Страница 12: ...s have been correctly positioned press E to return to the main menu If the application requires flow rate to be the primary display i e shown on the large display the d P function enables the rate and...

Страница 13: ...ive access to the sub menu 5 9 1 AC rejection FrE WARNING If the mains filter frequency is changed the rate display and the internal references if used must be recalibrated To provide maximum low freq...

Страница 14: ...ress E to return to the main menu 5 12 Function of the P push button C P This parameter defines the function of the P push button when the BA558C is in the display mode Either the input current in mil...

Страница 15: ...lect CLIP OFF from the main menu and press P to reveal the current setting This clip off threshold is shown as a percentage of total flow and may be adjusted in one percent increments using the Up or...

Страница 16: ...ection 7 1 describes the preferred method of calibration using an external traceable 4 20mA calibrator Section 7 2 contains an example of calibration using the BA558C internal 4 and 20mA references wh...

Страница 17: ...ress the Up or Down push button until the second least significant digit and following decimal point of the total display flash This gives the required total display resolution of 0 1 Press P to move...

Страница 18: ...to the main menu by pressing E Step 12 Reset the grand total to zero To reset the grand total to zero select CLr Gtot from the main menu and press P which will cause the instrument to display Clr no...

Страница 19: ...nd by displaying root which is the first function in the main menu Steps Perform as described in the 3 to 7 previous calibration example see section 6 1 Step 8 Calibrate the rate display Scroll throug...

Страница 20: ...e display 4 20mA current is noisy or the flow signal is unstable Eliminate source of electrical noise Reduce resolution of the rate display to mask flow noise Incorrect calibration Digital filter FrE...

Страница 21: ...ilable for a rate alarm and shows which are fail safe i e output is in the alarm condition open when the 4 20mA input current is zero Similarly when programmed as a total alarm the output should be op...

Страница 22: ...es the alarm function without changing the alarm parameters See section 8 3 3 tYPE Type of alarm Defines whether the alarm operates on the rate or total display See section 8 3 4 SPr1 Alarm setpoint 1...

Страница 23: ...YPE This function conditions the alarm to operate on the rate or total display Alarm 1 and Alarm 2 may both be conditioned as rate or total alarms or one may be conditioned for rate and the other for...

Страница 24: ...ing figure Each digit can be adjusted using the Up and Down push buttons and the P button to move to the next digit When the required hysteresis has been entered press E to return to the alarm menu e...

Страница 25: ...etpoints by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with this function disabled and the security code set to 0000 8 3 12 Adjusting alarm setpoints from th...

Страница 26: ...current limiting resistor in series with the supply will reduce the display brilliance 8 5 Additional front panel protection Although the BA558C has an IP65 sealed front panel for installations in hos...

Страница 27: ...otal 5 6 d P 5 3 5 6 EnbL 8 3 2 8 3 3 Fault finding during commissioning 7 1 after commissioning 7 2 Flow indicator 2 2 Flow transmitter loops 3 1 Grand total 2 1 clear 5 3 5 10 Guarantee 7 5 HI LO 8...

Отзывы: